4343W/4336W-G5/2530W-G5 X-ray Detectors
Software may be used to detect either of these sensors as 0 or 1 in various locations in an OEM
system, e.g. a wall stand, table bucky, or if a grid is in use.
OEM Sensor Recommendations:
1
The magnets used in the OEM bucky or grid should have a magnetic strength
rating between
1.17T and 1.21T
. This will provide a magnetic field to activate the sensor in
the detector without activating both of them, if they are appropriately placed.
2
The magnets used in the OEM bucky or grid must be oriented in a way that the polar fields
match.
3
The magnet dimensions used by the OEM should be approximately a
10mm x 1mm disk
.
13.5
Autonomous Mode
The Autonomous Mode vTrigger feature allows the wireless detector to run in an autonomous
mode where it can be armed, is then disconnected from the wireless access point while remain-
ing armed and ready for AED (Automatic Exposure Detection). In this mode, X-ray images are
stored on the detector to be transferred to a client PC later.
The detector can be armed in the autonomous mode and then taken to the patient's room. In the
process, the detector link will be closed and remain in the armed state, ready for X-ray. After one
or more images are acquired, the detector can be synchronized again with a client PC and
image(s) transferred so they can be associated with the patient study.
Refer to
document 06104
for details about the software interface with the Autonomous Mode
vTrigger operation.
WARNING
The OEM is responsible for synchronizing the images with the
respective patient names. Not complying with this requirement may
result in images associated with the wrong patient.
Note
This feature does not record names or personal patient information. Varex
only provides numerically named images. Personal information must be
managed by the OEM's software.
13.6
Detector Sharing
The X-ray detector is equipped with Detector Sharing capability. Detector Sharing allows the IP
address, password, and SSID to be set via software command. The software command for
Detector Sharing can be sent via wireless or the Service Cable. Refer to document
03236 FP-2032 - Varex Smart Panel Software Interface
for more information.