2315NDT v2 CMOS X-ray Detector User Manual
shows a schematic example of this trigger.
lists the timing values.
The falling edge of the SyncOut 2 signal is designed to facilitate the synchronization between
detector and generator as it is coupled with the end of the readout cycle.
Figure 12
Continuous readout - Duration Trigger Schematic
Table 14
Duration Trigger specification for Continuous readout
Symbol
Description
Value
Notes
T
D
Delay time
1.1 ms.
N/A
T
R/O
Frame readout
time
See frame readout time in
.
N/A
T
BR
Before readout
time
Width = 2.1 ms.
N/A
T
R
Detector reset
time
Minimum = 200
µs
Maximum = See Frame readout
times in
T
R
is measured from the rising
edge of the SyncIn to the end of
T
BR
+T
D
T
XRAY
Duration of the X-
ray pulse
Set by the user.
N/A