USER
MANUAL
20
www.vareximaging.com
2315NDT v2 CMOS X-ray Detector User Manual
•
Auto-descramble
•
Dark correction
•
Gain correction
•
Defect correction
2
Place a phantom or a test object on the detector surface, within the FoV
3
With the X-ray beam on, capture a single image by clicking
Snap
.
4
The result should be a clean-looking image with no obvious background shading, striping,
artifacts, or lines, as shown in
.
Figure 5
Sample Image
Note
If the detector is moved or other X-ray parameters are needed, the calibra-
tion must be performed again.
Note
It is important that the detector is not moved once calibrated and that the
exact same X-ray parameters used for the Flood field correction are used for
the final corrected X-ray image.