Trouble
Fault State
Possible
Cause
Method
There is
condensation
in the PF
water trap
Insufficient
pressure
Internal
pressure
gauge below
recommended
operating
levels
Call for service
Airflow is
blocked
Air inlet or
exhaust are
blocked or air
filter needs
cleaning
Clean air filter
and ensure that
the air inlets are
not blocked
Cooling fan is
not operating
(no airflow from
exhaust port
on bottom of
compressor)
The fan has
failed
Call for service
Air flow
required of the
compressor
exceeds
capabilities.
FiO
2
and flow
rates exceed
the guidance
for high altitude
operation
provided on
page 8
Follow
guidance on
page 8
Water trap
is collecting
moisture not
removed within
the compressor
Operation at
high altitude or
high flow rates
Condensation
is normal.
Check the
water bottle
reservoir
every 8 hours
and empty as
needed.
Page 28
Disposal
The Q50 Compressor should be disposed of in accordance with the
user’s defined disposal procedure for electromechanical equipment.