Trouble

Fault State

Possible 

Cause

Method

There is 

condensation 

in the PF 

water trap

Insufficient 

pressure

Internal 

pressure 

gauge below 

recommended 

operating 

levels

Call for service

Airflow is 

blocked

Air inlet or 

exhaust are 

blocked or air 

filter needs 

cleaning

Clean air filter 

and ensure that 

the air inlets are 

not blocked

Cooling fan is 

not operating 

(no airflow from 

exhaust port 

on bottom of 

compressor)

The fan has 

failed

Call for service

Air flow 

required of the 

compressor 

exceeds 

capabilities.

FiO

2

 and flow 

rates exceed 

the guidance 

for high altitude 

operation 

provided on 

page 8

Follow 

guidance on 

page 8

Water trap 

is collecting 

moisture not 

removed within 

the compressor

Operation at 

high altitude or 

high flow rates

Condensation 

is normal. 

Check the 

water bottle 

reservoir 

every 8 hours 

and empty as 

needed.

Page 28

Disposal

The Q50 Compressor should be disposed of in accordance with the 

user’s defined disposal procedure for electromechanical equipment.

Содержание Q50

Страница 1: ...Q50 Compressor Instructions for Use...

Страница 2: ......

Страница 3: ...and Weight 7 High Altitude Considerations 7 Working Principle and Construction 9 Working Principle 9 System Overview 9 Set Up for Use 13 Mounting Positions 13 Pole Mount 14 Install High Altitude High...

Страница 4: ...oxygen supply to the DISS oxygen fitting on the oxygen filter assembly on the back of the Precision Flow if oxygen concentrations above 21 are desired and pressurize the oxygen source 7 Attach the Q50...

Страница 5: ...lied air The compressor was designed to be mounted on a Precision Flow Roll Stand or placed directly on a flat level floor that does not obstruct the air exhaust underneath the compressor The low prof...

Страница 6: ...ould be used ONLY by a trained operator Contraindications The Compressor is not intended for use in MRI environments For more information see the Precision Flow Instructions for Use Warnings and Cauti...

Страница 7: ...de High Flow Kit Auto draining water trap assembly Water Bottle Water Bottle Holder Drain Tube 1 Pole mounts and one air intake filter are installed during manufacturing Related Documents Warranty Cer...

Страница 8: ...IEC 60601 1 medical electrical equipment Part 1 general safety requirement Environmental Requirements Relative Temperature Humidity Operation 18 30 C 10 90 Storage 10 50 C 20 90 CAUTION If the storag...

Страница 9: ...level conditions The density of air is lower at higher altitudes and has an impact on the flow capacity of the compressor The following table provides guidance for operating at higher altitudes CAUTI...

Страница 10: ...1 31 40 40 40 40 40 39 37 34 32 32 40 40 40 40 40 39 37 34 32 33 40 40 40 40 40 40 38 35 33 34 40 40 40 40 40 40 38 35 33 35 40 40 40 40 40 40 38 36 34 36 40 40 40 40 40 40 39 36 34 37 40 40 40 40 40...

Страница 11: ...he heat exchange coil and into the compressor Note Excess moisture is directed to the water bottle reservoir Here it is compressed moves through the heat exchange coil in order to reduce the air tempe...

Страница 12: ...Intake Filter 2 Power Inlet Switch and Fuse Drawer 3 Hour Meter 4 Serial Number Label 5 DISS Air Fitting Figure 2 Air Intake Filter Power Inlet Switch and Fuse Drawer Hour Meter DISS Air Fitting Seria...

Страница 13: ...rth Air Intake The compressor utilizes air from its surrounding environment It passes through multiple stages of filtration prior to delivery to the Precision Flow The first filtration stage is access...

Страница 14: ...compressor This could result in failure to deliver therapy as intended When the unit is placed on the floor be sure that any flooring such as carpet does not restrict the air exhaust WARNING Do not re...

Страница 15: ...compressor the correct distance from the floor to accommodate cooling air to exhaust from the unit WARNING Do not remove the compressor feet These are designed to reduce noise and provide the require...

Страница 16: ...the compressor must be placed on the Precision Flow Roll Stand base and no greater than 8 from the ground The compressor also must be placed over two locking casters of the Precision Flow Roll Stand I...

Страница 17: ...right side place the Vapotherm compressor on the right of the roll stand base Note The side with the water bottle reservoir should face front 2 PlaceonehalfofapoleclampbetweentheVapothermcompressor mo...

Страница 18: ...this ring will release the fitting on the water trap Pull the water trap away from the Precision Flow 2 Install the drain tube into the fitting on the bottom of the water trap Gently pull on the tube...

Страница 19: ...e end of the air hose onto the DISS air fitting on the back of the Q50 Compressor until it reaches the threads Rotate the collar on the air hose clockwise so that the threads engage with the DISS air...

Страница 20: ...or and Precision Flow unit before turning the unit on 4 Connect oxygen supply to Precision Flow if applicable Connect an oxygen supply as directed in the Precision Flow Instructions for Use if oxygen...

Страница 21: ...s been properly connected to the Precision Flow unit prior to powering on the compressor WARNING Ensure the Precision Flow is powered off before starting the compressor After powering on the compresso...

Страница 22: ...drain tube into the cap of the water bottle and pressing the water bottle into the water bottle holder NOTE It is normal for some water to remain in the water trap bowl while draining the majority of...

Страница 23: ...nic emissions IEC 61000 3 2 Not applicable The Vapotherm Q50 compressor is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power suppl...

Страница 24: ...water bottle reservoir and reservoir tubing according to hospital disinfection policies after emptying the water bottle reservoir and in between patient uses Do not allow liquids to infiltrate beyond...

Страница 25: ...l Water Trap and Air Hose Exterior Wipe the exterior according to hospital disinfection policies Use clean water to remove any residual residue Do not allow liquid to enter the outlet of the water tra...

Страница 26: ...g used CAUTION To prevent damage to the unit Always remove power cord prior to any cleaning or disinfection of the compressor If there are any questions regarding the safety or use of the disinfecting...

Страница 27: ...r 4 000 hours Air intake muffler replacement once per 4 000 hours Compressor rebuild once per 8 000 hours Compressor motor replacement once per 20 000 hours Replace the filter element Vapotherm Part N...

Страница 28: ...e Power cord is not connected or connected to a faulty outlet Connect power cord and ensure that the outlet is functioning Power cord is connected to a functioning power outlet but there is no audible...

Страница 29: ...an is running but not the compressor even when the compressor is cool The compressor has failed Call for service The compressor and cooling fan are running The air pipeline is not tightened properly T...

Страница 30: ...rt on bottom of compressor The fan has failed Call for service Air flow required of the compressor exceeds capabilities FiO2 and flow rates exceed the guidance for high altitude operation provided on...

Страница 31: ...paid by the Customer Customer shall also be responsible for the cost of labor for repairs The warranty set forth herein shall become null and void if 1 the Q50 device is not used or serviced in accord...

Страница 32: ...100 Domain Drive Exeter NH 03833 For customer service call 1 855 557 8276 Fax 603 218 6111 Email ts vtherm com Q50 Compressor Instructions for Use 3100772 Rev E...

Отзывы: