4. Repeat for the lower pole mount.
5. Firmly tighten both pole mounts.
Install High Altitude / High Flow Kit
1. Remove the manual draining water
trap from the air inlet to the PF. Using
a pair of needle nose pliers, push
the retaining ring of the air fitting on
the back of the Precision Flow unit in
towards the Precision Flow. Pushing
in this ring will release the fitting on
the water trap. Pull the water trap
away from the Precision Flow.
2. Install the drain tube into the fitting on
the bottom of the water trap. Gently
pull on the tube to confirm that it is
captured by the fitting.
3. Install the Auto draining water trap
into the Air inlet of the Precision Flow
by pressing the fitting on the water
trap fully into the air fitting on the
back of the Precision Flow. To ensure
that the fitting have completely
engaged, gently pull on the water
trap.
Page 16