thermistor that will automatically discontinue operation of the
compressor in the event of an overheating event.
Due to the inherent design of the Q50 compressor, while the unit is
running, any air not demanded by the Precision Flow unit is simply
recirculated within the system. This allows the compressor to run
continuously without generating excessive pressure within the
system or delivering flow not needed by the Precision Flow.
The water bottle reservoir should be checked once a shift (or every
8 hours) and emptied as appropriate. Actual emptying requirements
will depend on environmental conditions.
To empty the water bottle:
1. Pull the water bottle from the water bottle holder and slide the
draining tube out of the water bottle cap.
2. Empty the contents of the water bottle.
3. Reinstall by pushing the drain tube into the cap of the water
bottle and pressing the water bottle into the water bottle holder.
NOTE:
It is normal for some water to remain in the water
trap bowl while draining the majority of the water to the water
bottle.
When emptying the bottle, the water trap may also be emptied if
desired.
To empty the water trap:
1. While the compressor is running, gently turn the gray knob
on the bottom of the water trap towards the “O” marking. This
manually opens the drain on the water trap.
2. When the water trap is empty, gently turn the gray knob in the
direction of the “S” marking. This engages the seals within the
water trap. Do not over tighten the gray knob.
CAUTION:
Using tools to tighten the gray knob on the
bottom of the water trap may result in overtightening
and damage to the High Altitude / High Flow Kit.
Shutting down the system
In order to shut the compressor system down, simply press the
power inlet switch in the rear of the unit to the “ O ” position and the
compressor will cease operation.
Page 20