General Indications & Contraindications.

Primary Indications:

Precision Flow® is intended for use to add warm moisture to breathing gases from an external 
source for administration to a neonate/infant, pediatric and adult patients in the hospital, 
subacute institutions, and home settings. It adds heat and moisture to a blended medical air/ 
oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral 
oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

Contraindications:

General:

 

Any situations in which humidification is contra-indicated (see American Association of 
Respiratory Care Clinical Practice Guidelines).

 

Specific to Nasal Cannula: 
Patients with occluded or defective nares should not use the system.

Warnings and Cautions 

Warning

 indicates that a situation may occur which is potentially harmful to the patient or user.

Caution

 indicates a condition that may lead to equipment damage, malfunction, or 

inaccurate operation. A Note indicates a point of emphasis to make operation more efficient or 
convenient. 

Please take the time to familiarize yourself with the warnings, cautions, and notes listed in this 
manual. They cover safety considerations, special requirements, and regulations. 

The user of this product shall have sole responsibility for any malfunction due to operation 
or maintenance performed by anyone not trained by Vapotherm staff or official training 
documentation.
When handling any part of the Precision Flow®, always follow hospital infection control 
guidelines and Standard Precautions. Vapotherm also recommends that users follow the 
Centers for Disease Control (CDC) publications: Guidelines for Maintenance of In-Use 
Respiratory Therapy Equipment and Guidelines for Prevention of Nosocomial Pneumonia.

General Warnings

Federal Law (U.S.) restricts the sale of this device to, or by the order of any physician. This 
device should be used ONLY by a trained operator.
This is a humidification device generally used for providing continuous flows of breathing gas.The 
Precision Flow™ is not a ventilator and should not be used as life support.
Oxygen supports combustion; this device should not be used near or around open flames, oil, or 
grease, or flammables.
Service on the device should only be performed by qualified, certified service technicians.
To prevent injury, do not attempt to do any service to the Precision Flow® while a patient is 
connected to the device.
If the device is damaged or not working properly, do not use. Contact Vapotherm or your 
authorized Vapotherm representative.
Do not operate if power cord is damaged.
The device should not be turned on and left unattended if not on a patient. 
Do not use the Precision Flow® in or around water, other than the water bag that feeds the 
system.
Prior to use, the Precision Flow® should be positioned and secured to a sturdy IV pole with the 
base of the unit no more than 40” (102cm) above the floor to reduce risk of tipping.

Содержание precision flow

Страница 1: ...Precision Flow Operating Instruction Manual...

Страница 2: ...ction 2 Overview 6 Section 3 Principles of operation 7 Section 4 Controls displays connections 8 Section 5 Modes of operation 11 Section 6 Initial assembly 12 Section 7 Setting up 13 Section 8 Adjustm...

Страница 3: ...n Emerging and Newly Identified Health Risks SCENIHR has evaluated the exposure to DEHP for the general population and patients during medical procedures In some cases the exposure is significant and...

Страница 4: ...ion or maintenance performed by anyone not trained by Vapotherm staff or official training documentation When handling any part of the Precision Flow always follow hospital infection control guideline...

Страница 5: ...ounding connection do not operate the device Medical electrical equipment needs special precautions regarding electromagnetic radiation Portable and mobile RF communications equipment can affect medic...

Страница 6: ...e adjusted via a single setting control knob on the front panel All values and alarms displayed in a single large color coded panel Flow range 1 40 lpm Oxygen percentage is fully adjustable from 21 to...

Страница 7: ...charged The battery is not intended for patient transport gas flow will cease Disposable patient circuit The disposable patient circuit DPC is comprised of the disposable water path DWP vapor transfer...

Страница 8: ...ge type 4 Vapor transfer cartridge fault 5 Gas supply fault 6 Status LED 7 Run Standby button see note 8 Setting control knob 9 Alarm mute button 10 Alarm muted LED 11 General fault 12 Water out 13 Bl...

Страница 9: ...LED indicates one or more alarms are muted 4 Setting control knob Press to select which variable to adjust Rotate to adjust to new value Press again to set value 5 Hinged door Opens to install or remo...

Страница 10: ...nt 3 Access panel for oxygen sensor see note 4 Pole clamp 5 Power cord connection and fuse holder 6 DISS or NIST oxygen connection 7 DISS or NIST air connection 8 Gas inlet filters and traps Note Writ...

Страница 11: ...scrub the ports Do not apply organic solvents or bleach Section 5 Modes of operation Mode Action Indicator light color Sleep Display in sleep mode no gas flow Amber Standby Input parameters can be ad...

Страница 12: ...installed before first time use The filter and trap assemblies have a quick disconnect fitting which connects to the main unit and a DISS gas fitting for either an oxygen or an air hose Note The quic...

Страница 13: ...t not more than 40 102cm above the floor Fixed rail supports may also be used Use with Vapotherm approved IV poles 7 4 Connect oxygen and air supply hoses to correct inlets then connect them to the wa...

Страница 14: ...5 2 Fit the delivery tube to the disposable water path as shown Press firmly into place Insert fully Both latches must click shut Assembled disposable patient circuit ready for insertion Water inlet...

Страница 15: ...ward as shown 7 6 3 Slide disposable patient circuit downward into the docking station until it stops 7 6 4 Press down firmly to ensure correct seating NOTE If the hinged door does not close easily ch...

Страница 16: ...emperature There will also be a corresponding cartridge indicator on the lower right hand side which will identify what type of disposable patient circuit is in place Blue High or Red Low To enter Run...

Страница 17: ...tube to stop the flow of water into disposable patient circuit whenever the unit is in standby mode To adjust settings See section 8 Adjustments For alarms and troubleshooting See section 12 Alarms S...

Страница 18: ...down unit returns to default settings Section 9 Connecting to patient 9 1 Wait for desired set temperature to be reached before placing the cannula on the end of the patient delivery tube The flashin...

Страница 19: ...s at flow rates less than 5 lpm low flow cartridge or less than 10 lpm high flow cartridge To minimize condensation it is recommended not to set the temperature higher than 34 C if using flow rates le...

Страница 20: ...es 11 8 Slide the disposable patient circuit into the docking station and close the door 11 9 Hang new sterile water on the IV pole hook 11 10 Wipe the spike on the water inlet tube with 70 90 isoprop...

Страница 21: ...E PRIORITIES MEDIUM PRIORITY alarms require immediate attention and are indicated by rapid intermittent tones fast triple beeps LOW PRIORITY alarms require attention as soon as reasonably possible and...

Страница 22: ...cannula type re install DPC Water out flashing Medium Priority No water in disposable water path Gas flow continues without heat ing or water circulation Sterile water empty or obstructed inlet tube D...

Страница 23: ...path cartridge delivery tube None Cartridge and or DPC not detected STANDBY mode missing cartridge Remove disposable patient circuit Check cartridge installation Cartridge type None Indicates type of...

Страница 24: ...o 3003011 should be replaced annually It is accessible by removing a panel at the back of the unit and can be changed in a few minutes by the user or biomedical engineer Use only Vapotherm approved pa...

Страница 25: ...orite or 6 maximum Hydrogen Peroxide cleaning solution In addition the following detergent wipes may be used to remove any soil from the unit CaviwipesTM or Sani ClothTM AF3 Germicidal Unplug the Prec...

Страница 26: ...state depends on flow rate and temperature Back up power Qty 4 4 8V nickel metal hydride AA batteries not user replaceable Gas supply Medical air and oxygen at inlet pressures between 4 and 85 psi 28...

Страница 27: ...ISO 11195 ISTA 2A ENVIRONMENTAL Operating Ambient temperature 18 30 C Ambient relative humidity 0 90 RH non condensing Ambient Pressure Standard atmospheric Not to be used in hyperbaric conditions Sto...

Страница 28: ...uter Dia mm Max Flow Premature MN1100A 1 5 8 Neonatal MN1100B 1 5 8 Infant MI1300 1 9 8 Intermediate Infant MI1300B 1 9 8 SOLO cannula SOLO1300 1 9 8 Pediatric Small MPS1500 1 9 20 Pediatric Adult Sma...

Страница 29: ...on Flow enters RUN mode from STANDBY when the RUN STANDBY button is pressed Normal operation starts The pump heater and gas flow proportioning systems start Sensors and alarms are active and flow temp...

Страница 30: ...ation Power Frequency Magnetics EN 61000 4 8 1993 A1 2001 3A m 50 60Hz PASSED Voltage Dips and Dropouts EN 61000 4 11 2004 Per Standard The Precision Flow uses RF energy only for its internal function...

Страница 31: ...rized service center Labor for repairs will not be covered The warranty set forth in this shall become null and void if the Product is opened otherwise tampered with or if repairs are attempted by any...

Страница 32: ...Page 32 3001002 Rev P 3001002 Rev P Vapotherm Inc 22 Industrial Drive Exeter NH 03833 USA Phone 603 658 0011 Fax 603 658 0181...

Отзывы: