3001002 Rev. P    

 

Page  5

Section 1

  

 Indications, Warnings and Cautions

Make sure all Disposable Patient Circuit connections have been properly secured. 

The cartridge, disposable water path and delivery tube are labeled as 

single patient use

 only and 

must be replaced after 30 days use on a single patient: do not attempt to sterilize or reuse and 
follow all local and federal regulations for disposal. Outside the USA follow national or international 
regulations.

Failure to utilize sterile water supply or clean gas supply may increase risk of bacterial contamination.
  • Use aseptic technique.

  • Gas supply must be clean dry medical grade gas to prevent harm to the patient and prevent

    damage to the Precision Flow®

Precision Flow

®

 

is not a Continuous Positive Airway Pressure (CPAP) device.

 There are no 

controls to deliver or monitor airway pressure. Precision Flow® should not be used to deliver 
pressure in a closed system. 
Never connect the unit to a patient until it reaches set point temperature (temperature display stops 
flashing). Allow the unit to warm-up to purge condensate and prevent patient discomfort due to cold 
or partly humidified gas.

Additional patient monitoring is necessary if the Precision Flow® is used to give supplementary oxygen.

The Precision Flow

®

 is not MRI compatible.

The unit is provided with a Hospital Grade power cord. Do not use any other cord. 

Do not use 

extension cords

. For grounding reliability, the cord 

must

 be connected to an equivalent receptacle 

marked “Hospital Grade” or “Hospital Only”. If any doubt exists as to the grounding connection, 

do 

not 

operate the device.

Medical electrical equipment needs special precautions regarding electromagnetic radiation. Portable 
and mobile RF communications equipment can affect medical equipment and should not be used 
near the Precision Flow

®

The back-up battery is designed for temporary use only, when AC power to the unit has been 
interrupted. After the battery is fully discharged the device will not operate and patient gas flow will 
cease. There are no alarms or display indicators after the battery has discharged. The battery is not 
intended for patient transport.

General Cautions

Read and understand this manual prior to operating the system.
Clamp water supply when not in use, including Standby mode, to prevent damage by water ingress.

Aseptic techniques (including hand washing and avoiding touching connection points) and Standard 

Precautions should always be followed when handling medical equipment. Standard Precautions 

should always be followed when coming into contact with patients
Do not cover the unit; blocking the vent may damage the unit.

Do not:

  • Immerse the Precision Flow

®

 in water.

  • Steam or gas sterilize the Precision Flow

®

.

  • Wipe with bleach.
Flexible sterile water bags are recommended. If rigid or semi-rigid bottles are used, a Vapotherm 
approved venting bottle cap must be used.

NOTE

: The Precision Flow

®

 may be operated with limited performance at gas inlet pressures as low 

as 4 psi (28 kPa). However, for the full specified range of gas flows and oxygen percentages, both 
gas inlet pressures must be 40 psi (276 kPa) or above. Precision Flow has not been tested for use 
in Field transport. When used with approved ancillary equipment the Precision Flow may be used for 
transferring patients within the hosiptal.

Содержание precision flow

Страница 1: ...Precision Flow Operating Instruction Manual...

Страница 2: ...ction 2 Overview 6 Section 3 Principles of operation 7 Section 4 Controls displays connections 8 Section 5 Modes of operation 11 Section 6 Initial assembly 12 Section 7 Setting up 13 Section 8 Adjustm...

Страница 3: ...n Emerging and Newly Identified Health Risks SCENIHR has evaluated the exposure to DEHP for the general population and patients during medical procedures In some cases the exposure is significant and...

Страница 4: ...ion or maintenance performed by anyone not trained by Vapotherm staff or official training documentation When handling any part of the Precision Flow always follow hospital infection control guideline...

Страница 5: ...ounding connection do not operate the device Medical electrical equipment needs special precautions regarding electromagnetic radiation Portable and mobile RF communications equipment can affect medic...

Страница 6: ...e adjusted via a single setting control knob on the front panel All values and alarms displayed in a single large color coded panel Flow range 1 40 lpm Oxygen percentage is fully adjustable from 21 to...

Страница 7: ...charged The battery is not intended for patient transport gas flow will cease Disposable patient circuit The disposable patient circuit DPC is comprised of the disposable water path DWP vapor transfer...

Страница 8: ...ge type 4 Vapor transfer cartridge fault 5 Gas supply fault 6 Status LED 7 Run Standby button see note 8 Setting control knob 9 Alarm mute button 10 Alarm muted LED 11 General fault 12 Water out 13 Bl...

Страница 9: ...LED indicates one or more alarms are muted 4 Setting control knob Press to select which variable to adjust Rotate to adjust to new value Press again to set value 5 Hinged door Opens to install or remo...

Страница 10: ...nt 3 Access panel for oxygen sensor see note 4 Pole clamp 5 Power cord connection and fuse holder 6 DISS or NIST oxygen connection 7 DISS or NIST air connection 8 Gas inlet filters and traps Note Writ...

Страница 11: ...scrub the ports Do not apply organic solvents or bleach Section 5 Modes of operation Mode Action Indicator light color Sleep Display in sleep mode no gas flow Amber Standby Input parameters can be ad...

Страница 12: ...installed before first time use The filter and trap assemblies have a quick disconnect fitting which connects to the main unit and a DISS gas fitting for either an oxygen or an air hose Note The quic...

Страница 13: ...t not more than 40 102cm above the floor Fixed rail supports may also be used Use with Vapotherm approved IV poles 7 4 Connect oxygen and air supply hoses to correct inlets then connect them to the wa...

Страница 14: ...5 2 Fit the delivery tube to the disposable water path as shown Press firmly into place Insert fully Both latches must click shut Assembled disposable patient circuit ready for insertion Water inlet...

Страница 15: ...ward as shown 7 6 3 Slide disposable patient circuit downward into the docking station until it stops 7 6 4 Press down firmly to ensure correct seating NOTE If the hinged door does not close easily ch...

Страница 16: ...emperature There will also be a corresponding cartridge indicator on the lower right hand side which will identify what type of disposable patient circuit is in place Blue High or Red Low To enter Run...

Страница 17: ...tube to stop the flow of water into disposable patient circuit whenever the unit is in standby mode To adjust settings See section 8 Adjustments For alarms and troubleshooting See section 12 Alarms S...

Страница 18: ...down unit returns to default settings Section 9 Connecting to patient 9 1 Wait for desired set temperature to be reached before placing the cannula on the end of the patient delivery tube The flashin...

Страница 19: ...s at flow rates less than 5 lpm low flow cartridge or less than 10 lpm high flow cartridge To minimize condensation it is recommended not to set the temperature higher than 34 C if using flow rates le...

Страница 20: ...es 11 8 Slide the disposable patient circuit into the docking station and close the door 11 9 Hang new sterile water on the IV pole hook 11 10 Wipe the spike on the water inlet tube with 70 90 isoprop...

Страница 21: ...E PRIORITIES MEDIUM PRIORITY alarms require immediate attention and are indicated by rapid intermittent tones fast triple beeps LOW PRIORITY alarms require attention as soon as reasonably possible and...

Страница 22: ...cannula type re install DPC Water out flashing Medium Priority No water in disposable water path Gas flow continues without heat ing or water circulation Sterile water empty or obstructed inlet tube D...

Страница 23: ...path cartridge delivery tube None Cartridge and or DPC not detected STANDBY mode missing cartridge Remove disposable patient circuit Check cartridge installation Cartridge type None Indicates type of...

Страница 24: ...o 3003011 should be replaced annually It is accessible by removing a panel at the back of the unit and can be changed in a few minutes by the user or biomedical engineer Use only Vapotherm approved pa...

Страница 25: ...orite or 6 maximum Hydrogen Peroxide cleaning solution In addition the following detergent wipes may be used to remove any soil from the unit CaviwipesTM or Sani ClothTM AF3 Germicidal Unplug the Prec...

Страница 26: ...state depends on flow rate and temperature Back up power Qty 4 4 8V nickel metal hydride AA batteries not user replaceable Gas supply Medical air and oxygen at inlet pressures between 4 and 85 psi 28...

Страница 27: ...ISO 11195 ISTA 2A ENVIRONMENTAL Operating Ambient temperature 18 30 C Ambient relative humidity 0 90 RH non condensing Ambient Pressure Standard atmospheric Not to be used in hyperbaric conditions Sto...

Страница 28: ...uter Dia mm Max Flow Premature MN1100A 1 5 8 Neonatal MN1100B 1 5 8 Infant MI1300 1 9 8 Intermediate Infant MI1300B 1 9 8 SOLO cannula SOLO1300 1 9 8 Pediatric Small MPS1500 1 9 20 Pediatric Adult Sma...

Страница 29: ...on Flow enters RUN mode from STANDBY when the RUN STANDBY button is pressed Normal operation starts The pump heater and gas flow proportioning systems start Sensors and alarms are active and flow temp...

Страница 30: ...ation Power Frequency Magnetics EN 61000 4 8 1993 A1 2001 3A m 50 60Hz PASSED Voltage Dips and Dropouts EN 61000 4 11 2004 Per Standard The Precision Flow uses RF energy only for its internal function...

Страница 31: ...rized service center Labor for repairs will not be covered The warranty set forth in this shall become null and void if the Product is opened otherwise tampered with or if repairs are attempted by any...

Страница 32: ...Page 32 3001002 Rev P 3001002 Rev P Vapotherm Inc 22 Industrial Drive Exeter NH 03833 USA Phone 603 658 0011 Fax 603 658 0181...

Отзывы: