background image

3001002 Rev. P    

 

Page  7

Section 3   

Principals of operation

The Precision Flow

®

 warms and humidifies breathing gas for delivery by nasal cannula at flows 

from 1 to 40 lpm. The unit incorporates an electronic blender and flow sensors that allow the 

oxygen percentage and total gas flow to be set independently.
The Precision Flow

® 

 consists of two parts:

Main unit

• The 

main unit

 which contains all the electrical and electronic components including the 

electronic blender and flow controllers, and remote sensors to monitor the disposable water 

path. The main unit has no water pathways and the gas pathway contains only dry gas at room 

temperature, and consequently does not need internal cleaning or disinfection.

•  The flow of oxygen and air are measured by 

mass flow sensors

. The operating software 

calculates the required flow of each needed to reach the target flow and oxygen percentage 

set by the operator. The system controls gas flows accordingly by adjusting proportional 

solenoid valves

 on the gas lines. An 

oxygen sensor

 monitors the gas mixture and signals any 

discrepancy between target and measured percentage. The oxygen sensor is automatically 

calibrated with oxygen at power-up and every 24 hours.

• 

Firmware 

running in the main unit uses sensors to monitor gas pressure, water temperature, and 

to detect air leaks into the water circuit (bubble detector). Alarms are displayed if any parameters 

are outside the normal range. Other indicators show low charge in the backup battery, and the type 

of cartridge installed. See Appendix for a description of the firmware states and transitions.

•  After a two hour charging period, an internal

 battery

 backup will maintain the set flow and 

oxygen blend for at least 15 minutes without AC power. The battery is not operator replaceable.

WARNING

: The back-up battery is designed for temporary use only, when AC power to the unit 

has been interrupted. After the battery is fully discharged the device will not operate and patient 

gas flow will cease.  There are no alarms or display indicators after the battery has discharged. 

The battery is not intended for patient transport. gas flow will cease.  

Disposable patient circuit

•  The 

disposable patient circuit

 (DPC) is comprised of the disposable water path (DWP), vapor  

  transfer cartridge (VTC) and delivery tube. Conditions in the circulating water and gas streams  

  are sensed remotely via the interface between the main unit and the disposable water path.
• 

Vapor transfer cartridge

. In the cartridge, blended gas passes through the lumens of hundreds  

  of parallel hollow fibers made of a specially developed polymer. Warm water circulates around  

  the fibers and diffuses as vapor through the fiber material into the gas stream flowing through  

  each fiber. Unlike most humidifiers, there is no direct contact between the water and gas  

  streams. The gas stream leaves the cartridge saturated with vapor at the set temperature.
 

Note

: Use 

only

 approved cartridges from Vapotherm Inc.

• 

Patient delivery tube.

The warmed humidified gas passes through the center of a triple-lumen  

  heated delivery tube. The center lumen is surrounded by two outer lumens circulating warmed  

   water to maintain the temperature of the inner lumen and to minimize rain-out. A proprietary short  

  nasal cannula is connected to the end of the delivery tube and passes the humidified breathing  

  gas to the patient’s nares. It is normal for non-DEHP PVC tubing to appear slightly cloudy, or  

  yellow, especially during longer use or when operated at higher temperatures.
• 

Disposable water path.

 The disposable water path houses a water reservoir, pump,connections  

  for the cartridge and delivery tube, and sensor interfaces to the main unit. Water is pumped past  

  a heater plate through the outer lumens of the delivery tube. Returning water passes through the  

  outer jacket of the specially designed vapor transfer cartridge where some water is lost as  

  vapor to the gas stream. There is no direct contact between water and gas flows. The water  

  then returns to the pump reservoir. Heater power automatically maintains the set temperature.  

  Water flows into the circuit from the water bag to replace evaporative losses in the vapor transfer  
  cartridge. Air is purged to atmosphere from the circulation via a hydrophobic filter membrane. 

See Section 5 for a description of the modes of operation.

Содержание precision flow

Страница 1: ...Precision Flow Operating Instruction Manual...

Страница 2: ...ction 2 Overview 6 Section 3 Principles of operation 7 Section 4 Controls displays connections 8 Section 5 Modes of operation 11 Section 6 Initial assembly 12 Section 7 Setting up 13 Section 8 Adjustm...

Страница 3: ...n Emerging and Newly Identified Health Risks SCENIHR has evaluated the exposure to DEHP for the general population and patients during medical procedures In some cases the exposure is significant and...

Страница 4: ...ion or maintenance performed by anyone not trained by Vapotherm staff or official training documentation When handling any part of the Precision Flow always follow hospital infection control guideline...

Страница 5: ...ounding connection do not operate the device Medical electrical equipment needs special precautions regarding electromagnetic radiation Portable and mobile RF communications equipment can affect medic...

Страница 6: ...e adjusted via a single setting control knob on the front panel All values and alarms displayed in a single large color coded panel Flow range 1 40 lpm Oxygen percentage is fully adjustable from 21 to...

Страница 7: ...charged The battery is not intended for patient transport gas flow will cease Disposable patient circuit The disposable patient circuit DPC is comprised of the disposable water path DWP vapor transfer...

Страница 8: ...ge type 4 Vapor transfer cartridge fault 5 Gas supply fault 6 Status LED 7 Run Standby button see note 8 Setting control knob 9 Alarm mute button 10 Alarm muted LED 11 General fault 12 Water out 13 Bl...

Страница 9: ...LED indicates one or more alarms are muted 4 Setting control knob Press to select which variable to adjust Rotate to adjust to new value Press again to set value 5 Hinged door Opens to install or remo...

Страница 10: ...nt 3 Access panel for oxygen sensor see note 4 Pole clamp 5 Power cord connection and fuse holder 6 DISS or NIST oxygen connection 7 DISS or NIST air connection 8 Gas inlet filters and traps Note Writ...

Страница 11: ...scrub the ports Do not apply organic solvents or bleach Section 5 Modes of operation Mode Action Indicator light color Sleep Display in sleep mode no gas flow Amber Standby Input parameters can be ad...

Страница 12: ...installed before first time use The filter and trap assemblies have a quick disconnect fitting which connects to the main unit and a DISS gas fitting for either an oxygen or an air hose Note The quic...

Страница 13: ...t not more than 40 102cm above the floor Fixed rail supports may also be used Use with Vapotherm approved IV poles 7 4 Connect oxygen and air supply hoses to correct inlets then connect them to the wa...

Страница 14: ...5 2 Fit the delivery tube to the disposable water path as shown Press firmly into place Insert fully Both latches must click shut Assembled disposable patient circuit ready for insertion Water inlet...

Страница 15: ...ward as shown 7 6 3 Slide disposable patient circuit downward into the docking station until it stops 7 6 4 Press down firmly to ensure correct seating NOTE If the hinged door does not close easily ch...

Страница 16: ...emperature There will also be a corresponding cartridge indicator on the lower right hand side which will identify what type of disposable patient circuit is in place Blue High or Red Low To enter Run...

Страница 17: ...tube to stop the flow of water into disposable patient circuit whenever the unit is in standby mode To adjust settings See section 8 Adjustments For alarms and troubleshooting See section 12 Alarms S...

Страница 18: ...down unit returns to default settings Section 9 Connecting to patient 9 1 Wait for desired set temperature to be reached before placing the cannula on the end of the patient delivery tube The flashin...

Страница 19: ...s at flow rates less than 5 lpm low flow cartridge or less than 10 lpm high flow cartridge To minimize condensation it is recommended not to set the temperature higher than 34 C if using flow rates le...

Страница 20: ...es 11 8 Slide the disposable patient circuit into the docking station and close the door 11 9 Hang new sterile water on the IV pole hook 11 10 Wipe the spike on the water inlet tube with 70 90 isoprop...

Страница 21: ...E PRIORITIES MEDIUM PRIORITY alarms require immediate attention and are indicated by rapid intermittent tones fast triple beeps LOW PRIORITY alarms require attention as soon as reasonably possible and...

Страница 22: ...cannula type re install DPC Water out flashing Medium Priority No water in disposable water path Gas flow continues without heat ing or water circulation Sterile water empty or obstructed inlet tube D...

Страница 23: ...path cartridge delivery tube None Cartridge and or DPC not detected STANDBY mode missing cartridge Remove disposable patient circuit Check cartridge installation Cartridge type None Indicates type of...

Страница 24: ...o 3003011 should be replaced annually It is accessible by removing a panel at the back of the unit and can be changed in a few minutes by the user or biomedical engineer Use only Vapotherm approved pa...

Страница 25: ...orite or 6 maximum Hydrogen Peroxide cleaning solution In addition the following detergent wipes may be used to remove any soil from the unit CaviwipesTM or Sani ClothTM AF3 Germicidal Unplug the Prec...

Страница 26: ...state depends on flow rate and temperature Back up power Qty 4 4 8V nickel metal hydride AA batteries not user replaceable Gas supply Medical air and oxygen at inlet pressures between 4 and 85 psi 28...

Страница 27: ...ISO 11195 ISTA 2A ENVIRONMENTAL Operating Ambient temperature 18 30 C Ambient relative humidity 0 90 RH non condensing Ambient Pressure Standard atmospheric Not to be used in hyperbaric conditions Sto...

Страница 28: ...uter Dia mm Max Flow Premature MN1100A 1 5 8 Neonatal MN1100B 1 5 8 Infant MI1300 1 9 8 Intermediate Infant MI1300B 1 9 8 SOLO cannula SOLO1300 1 9 8 Pediatric Small MPS1500 1 9 20 Pediatric Adult Sma...

Страница 29: ...on Flow enters RUN mode from STANDBY when the RUN STANDBY button is pressed Normal operation starts The pump heater and gas flow proportioning systems start Sensors and alarms are active and flow temp...

Страница 30: ...ation Power Frequency Magnetics EN 61000 4 8 1993 A1 2001 3A m 50 60Hz PASSED Voltage Dips and Dropouts EN 61000 4 11 2004 Per Standard The Precision Flow uses RF energy only for its internal function...

Страница 31: ...rized service center Labor for repairs will not be covered The warranty set forth in this shall become null and void if the Product is opened otherwise tampered with or if repairs are attempted by any...

Страница 32: ...Page 32 3001002 Rev P 3001002 Rev P Vapotherm Inc 22 Industrial Drive Exeter NH 03833 USA Phone 603 658 0011 Fax 603 658 0181...

Отзывы: