Operating Instruction Manual
Page 17
Fig. 1
Section 8
Routine Disinfecting Protocol
8.1
Disinfection Supplies
1. DK-301 (a-e included in kit)
a. Disinfection Bag A
b. Disinfection Bag B
c. Disinfection Tube
d. Cartridge Bypass Tubes
e. Y-Spike Assembly
2. Gloves
3. Safety Glasses
4. Disinfectant wipes, 70-90% isopropyl alcohol
5. Approved Disinfectant
6. 1000ml Sterile Water Bag
7. Medical Grade Air Source
8. Standard adult flow meter with oxygen 7ft tubing attached
WARNINGS:
Vapotherm should be disinfected after each patient or every 30 days on a single patient. Do not disinfect in
an open patient care area.
The DK-301 disinfection kit is a single use item. Operators should open a new disinfection kit for each
disinfection procedure and discard the components at the end of the procedure.
Disinfection Procedure should be performed in a well ventilated area. Use Standard Precautions and
aseptic techniques during this procedure.
The Vapor Transfer Cartridge SHOULD NOT be in place when disinfecting the unit.
The Vapor Transfer Cartridge is a single use disposable and must be discarded after each patient use.
DK-301 IS NOT designed to disinfect Vapor Transfer Cartridges.
8.2
Pre-Cleaning Process
1. After patient use, it is recommended that the Vapotherm 2000i System remain attached to the
IV
pole with all the component parts intact.
2. Move the Vapotherm 2000i System to a hospital approved
reprocessing area outside the patient care area.
WARNING:
The water circuit of the Vapotherm 2000i System is not
sterile and can potentially have bacterial contamination. The water
circuit of the device should never be opened in a patient care area.
Transport the Vapotherm 2000i System to an appropriate area for
draining, cleaning and disinfection.
3. Wash hands and put on gloves.
4. Drain the Vapotherm 2000i System in a receptacle by cutting the
delivery tube.
(Fig. 1)
5. Remove and dispose of the delivery tube, cannula, VSS-1 spike
set, and sterile water source.
6. Remove and dispose of the vapor transfer cartridge.
WARNING:
The water circuit of the Vapotherm 2000i system has the potential for bacterial growth so
standard precautions should be used to open the water circuit. Disposable components should be disposed of
in accordance with hospital guidelines and operators should wash their hands after breaking down the device.
Operating Instruction Manual
Page 16
Fig. 1
Fig. 2
Section 7
Removing From Patient and System Shut Down
1. Remove cannula or other interface from patient.
2. Press and release power switch
(Fig. 1)
. Display will show “--”
3. Close clip on VSS-1
(Fig. 2)
.
NOTE:
The system’s pump continues to run for 1 minute to allow
heater to cool down.
4. After 1 minute, water pump shuts off and numeric display is
blank. Unit may now be disconnected from power outlet.
5. Remove unit from patient care area and proceed to disinfection
process.
CAUTIONS:
Avoid disconnecting from power or gas sources while machine is
operating.
Do not unplug from power source until display is blank.