Vapotherm 2000i Скачать руководство пользователя страница 8

Operating Instruction Manual 

Page 15

Fig. 1

Fig. 3

Section 6

 Alarms, Trouble Shooting and Component Change-Outs

6.3

  Component Change Outs

WARNINGS:

 

The vapor transfer cartridge, patient delivery tube, and VSS-1  
spike set are all single use only and should be discarded after  
removal from the Vapotherm 2000i.
The Patient Delivery Tube, VSS-1 and Vapor Transfer Cartridge  
should not be changed or replaced in the patient care area. 
The system must be disinfected any time the Vapor Transfer  
Cartridge, VSS-1 or Patient Delivery Tube are removed.

NOTE:

 The cannula and sterile water source can be replaced without  

disinfecting the system. As with all respiratory equipment, proper 
hand washing techniques should be followed before contacting 
or replacing any patient interfaces.

6.3.1

  Replacing Vapor Transfer Cartridge

1.  Power off unit. Disconnect gas flow.
2.  Close clip on VSS-1. 

(Fig. 1)

3.  Open hinged cover.
4.  Disconnect air tubes from cartridge ends by pressing tubing away 

 

from cartridge. 

5.  Remove cartridge by pulling straight outwards. 

(Fig. 2)

6.  Proceed to Section 8.0 and disinfect the Vapotherm

®

 2000i device 

 

before returning the device to service.  

7.  For set-up please refer to Section 4.3 of the manual.

CAUTION:

 Do not grip cartridge tubing with sharp instruments.

6.3.2

  Replacing the Patient Delivery Tube

1.  Power unit off. Disconnect gas flow.
2.  To remove tube, push base of tube upwards, rotate 1/4 turn 

 

counter clockwise and pull downward. 

(Fig. 3)

3.  Proceed to Section 8.0 and disinfect the Vapotherm

®

 2000i device 

 

before returning the device to service.

4.  For set-up please refer to Section 4.4 of the manual.

6.3.3

  Replacing the VSS-1 Spike Set

1.  Power off unit. Disconnect gas flow.
2.  Clamp VSS-1 and remove VSS-1 Spike Set from the water inlet

 

port on the back of the Vapotherm 2000i by releasing the

 

quick connect on the water inlet port.

3.  Proceed to Section 8.0 and disinfect the Vapotherm

®

 2000i device

 

before returning the device to service.

4.  For set-up please refer to Section 4.5 of the manual.

Fig. 2

Alarm 

Cause 

Action 

indication

Operating Instruction Manual 

Page 14

If further assistance is needed please call your clinical product specialist or local distributor representative.

S

S

If the Cartridge Alarm is continuous 

and air bubbles are rising into the 

VSS-1 bubble trap or if a flow or 
water is visible in the tube below 

the cartridge, then the cartridge 

has failed.

If cartridge alarm is intermittent 
and there are no bubbles in the  

VSS-1 bubble trap or no obvious 

water flow below the cartridge there 

may be condensation in the system.

First disconnect the patient from the unit, 

shut down unit, drain unit, disinfect unit, 

replace cartridge, VSS-1 and delivery tube, 
and follow set up instructions.

 

 

Occasional brief alarms due to 

condensation are not a cause for concern. 
Try briefly pinching and releasing tube 
under cartridge to dislodge the drops and/ 
or decrease set temperature.

Cartridge

Malfunction of Temperature 
Control System.

Shut down system and return for service. 

NOTE: 

A momentary High Temperature 

alarm may occur when the unit has 
been switched off and on again. If the 
temperature then stabilizes, no action 
is needed.

High 
Temperature 
Alarm

High water or air pressure due to 

high resistance in water circulation 
or air outlet: or malfunctioning 

pressure sensor.

Blocked tube alarm due to high 
WATER pressure will cause a 

continuous or intermittent tone 

and alarm light. The flow of 

breathing gas continues, but is 

no longer heated.

 

Blocked tube alarm due to high 
GAS pressure will cause a 5 second 
alarm tone. If the obstruction 

persists the system will continue  

to alarm in 5 sec episodes. Water 

circulation continues but the heater 

shuts off. 

 
 
 

Check that delivery tube is correctly 

positioned, rotated clockwise, and pulled 

into locked position. Check that water is 
circulating within delivery tube. If alarm 
persists replace delivery tube and/or 
cartridge. Disinfect unit prior to replacing 
components.

Find and correct the cause of obstruction.  
The most common cause is a kink in the 
nasal cannual or in the prong. Attempting 
to run the Vapotherm 2000i at very high 

flow through a patient interface not 

approved by Vapotherm may also raise the 

internal pressure sufficiently to trigger a 

Blocked Tube Alarm.

Blocked Tube 
Alarm

S

Section 6

 Alarms, Trouble Shooting and Component Change-Outs

Содержание 2000i

Страница 1: ...2000i Operating Instruction Manual...

Страница 2: ...his manual Vapotherm also recommends that users follow the Centers for Disease Control CDC publications Guidelines for Maintenance of In Use Respiratory Therapy Equipment and Guidelines for Prevention...

Страница 3: ...be used to deliver pressure in a closed system 2 3 Cautions Verify that the power source is compatible with the electrical specifications shown on each component For proper grounding reliability conne...

Страница 4: ...w flow cartridge VT01 BS Ensure that the correct cartridge is inserted before operating Section 3 About the Vapotherm 2000i The Vapotherm 2000i warms and humidifies flows of air oxygen or medical gas...

Страница 5: ...fer Cartridge 1 The Vapor Transfer Cartridge VT01 AS or VT01 BS attaches to the unit by two water and two air connections 2 When facing the unit access is via a hinged cover on right side The cartridg...

Страница 6: ...rtridge or Patient Delivery Tube and both should be checked or changed NOTE If using a low flow cartridge VT01 BS the flow cannot be decreased below 5 lpm until an appropriate cannula has been attache...

Страница 7: ...Check that water is properly circulating through the machine by making sure the Patient Delivery Tube is warm across the entire length 2 If good circulation cannot be confirmed check that the water f...

Страница 8: ...Manual Page 14 If further assistance is needed please call your clinical product specialist or local distributor representative S S If the Cartridge Alarm is continuous and air bubbles are rising int...

Страница 9: ...approved reprocessing area outside the patient care area Warning The water circuit of the Vapotherm 2000i System is not sterile and can potentially have bacterial contamination The water circuit of th...

Страница 10: ...70 90 isopropyl alcohol disinfectant wipes when wiping down the Vapotherm 2000i and disinfection kit components Always use a new disinfectant wipe taken directly from the package or container and ens...

Страница 11: ...ocol 8 5 Disinfect Gas and Water Circuits 1 Open clamp on Bag A Disinfectant will start to drain from Bag A through the unit and into Bag B Fig 9 2 When disinfectant has stopped draining into Bag B cl...

Страница 12: ...ral Information 9 1 Specifications Dimensions Height 11 280 mm width 5 5 140 mm depth 4 5 114 mm excluding IV pole clamp Weight Less than 6 lbs 2 7 kg without water reservoir Vapotherm spike set Works...

Страница 13: ...Appendix A WARNINGS These disinfectant solutions are designed to be used to disinfect the Vapotherm 2000i machine without the cartridge in place These disinfectant solutions ARE NOT approved to disinf...

Страница 14: ...syringe or pipette graduated in 1ml increments Concentrate Disinfectant Minncare Cold Sterilant EPA Reg No 52252 4 22 00 Hydrogen Peroxide 4 50 Peroxyacetic Acid 73 50 Inert Ingredients Control III Di...

Страница 15: ...Vapotherm Inc 198 Log Canoe Circle Stevensville MD 21666 T 001 410 604 3977 F 001 410 604 3978 www vtherm com PN 8 300068 00 Rev E bdc 8369 Supplied by...

Отзывы: