Vapotherm 2000i Скачать руководство пользователя страница 5

Operating Instruction Manual 

Page 9

Fig. 4

Fig. 5

Fig. 6

Section 4 

Set-up

4.5

  Vapotherm Spike Set (VSS-1): for  

connecting Sterile Water Bag

1.  Hang a sterile water bag from IV Pole.

2.  Connect VSS-1 to the water inlet port on back of unit and make 

 

sure it locks into place (Fig. 5).

3.  Ensure the VSS-1 is clamped then remove spike cap. Wipe spike 

 

with disinfectant wipes, 70-90% isopropyl alcohol.

4  Firmly insert spike into sterile water bag while avoiding direct 

 

hand contact with the spike tip and water bag septum.

5.  Leave VSS-1 spike set clamped until ready to fill unit.

WARNINGS: 

The VSS-1 is single patient use item and should be changed with 
each patient.

If the VSS-1 is removed from the Vapotherm device for any reason 
the Vapotherm 2000i should be disinfected following the routine  
disinfection procedure before being returned to service.

CAUTION:

 Never leave the VSS-1 unclamped when the system is not  

running.

NOTE: Removing an empty sterile water container does not constitute  
opening the closed system. New sterile water containers can be spiked  
using the same VSS-1 without removing the device from service  
following the procedure above.

4.6

  Connect To A Gas Source

1.  Connect a source of air, oxygen or medical gas blender to the gas 

 

inlet port of Vapotherm 2000i 

(Fig. 6)

. Gas inlet connection is a 

 

hose barb that accepts female fitting on a standard 1/4” 

 

(6.35mm) oxygen tube. 

NOTE: 

Vapotherm will not operate unless there is gas pressure at gas  

inlet. With no flow/pressure sensed, a “System Failure” alarm will  
sound.

Operating Instruction Manual 

Page 8

Fig. 1

Fig. 2

Fig. 3

Section 4 

Set-up

4.3

  Inserting the Vapor Transfer Cartridge

1.   The Vapor Transfer Cartridge (VT01-AS or VT01-BS) attaches to 

 

the unit by two water and two air connections.

2.   When facing the unit, access is via a hinged cover on right side. 

 

The cartridge may be fitted in either direction.

3.   Date the cartridge.

4.   Remove protective caps from luer side ports of cartridge

 (Fig. 1)

.

5.   Attach lower air tube from Vapotherm 2000i to lower end of 

 

cartridge.

6.   Insert projecting side ports into matching connections in unit.

 

Press cartridge firmly into place 

(Fig. 2)

.

7.  Attach upper air tube from Vapotherm 2000i to top of

 

cartridge 

(Fig. 3)

. Make sure tubing is not kinked.

8 .   Close hinged cover. If it does not close easily, check that cartridge 

 

is pressed fully into place and that air tubes are not interfering  

with cover.

WARNINGS:

 

The cartridge must be changed between patients and discarded 
after each use.

If the cartridge is removed, the unit should be disinfected.

If the cartridge is dropped, it should be discarded.

NOTE:

 Do not remove cartridge from the Vapotherm 2000i 

without first draining the machine.

4.4

  Inserting the Patient Delivery Tube

1.  Insert Patient Delivery Tube into lower portion of the unit by 

 

aligning blue tabs on tube with notches on bottom of unit.

2.  Firmly press into place 

(Fig. 4, see next page)

. Blue lip on tube 

 

must be flush with the bottom of unit.

3.  Rotate 1/4 turn clockwise and pull slightly downwards to lock into 

 

place (Fig. 4, 

see next page

).

WARNING:

 The Patient Delivery Tube is a single patient use item 

and should be changed with each patient.

If the Patient Delivery Tube is removed from the device for any reason  
the Vapotherm 2000i should be disinfected following the routine 
disinfection procedure before returning to service.

CAUTION: 

Unit will not operate correctly if the Patient Delivery Tube 

is inserted improperly or not locked into place. 

Содержание 2000i

Страница 1: ...2000i Operating Instruction Manual...

Страница 2: ...his manual Vapotherm also recommends that users follow the Centers for Disease Control CDC publications Guidelines for Maintenance of In Use Respiratory Therapy Equipment and Guidelines for Prevention...

Страница 3: ...be used to deliver pressure in a closed system 2 3 Cautions Verify that the power source is compatible with the electrical specifications shown on each component For proper grounding reliability conne...

Страница 4: ...w flow cartridge VT01 BS Ensure that the correct cartridge is inserted before operating Section 3 About the Vapotherm 2000i The Vapotherm 2000i warms and humidifies flows of air oxygen or medical gas...

Страница 5: ...fer Cartridge 1 The Vapor Transfer Cartridge VT01 AS or VT01 BS attaches to the unit by two water and two air connections 2 When facing the unit access is via a hinged cover on right side The cartridg...

Страница 6: ...rtridge or Patient Delivery Tube and both should be checked or changed NOTE If using a low flow cartridge VT01 BS the flow cannot be decreased below 5 lpm until an appropriate cannula has been attache...

Страница 7: ...Check that water is properly circulating through the machine by making sure the Patient Delivery Tube is warm across the entire length 2 If good circulation cannot be confirmed check that the water f...

Страница 8: ...Manual Page 14 If further assistance is needed please call your clinical product specialist or local distributor representative S S If the Cartridge Alarm is continuous and air bubbles are rising int...

Страница 9: ...approved reprocessing area outside the patient care area Warning The water circuit of the Vapotherm 2000i System is not sterile and can potentially have bacterial contamination The water circuit of th...

Страница 10: ...70 90 isopropyl alcohol disinfectant wipes when wiping down the Vapotherm 2000i and disinfection kit components Always use a new disinfectant wipe taken directly from the package or container and ens...

Страница 11: ...ocol 8 5 Disinfect Gas and Water Circuits 1 Open clamp on Bag A Disinfectant will start to drain from Bag A through the unit and into Bag B Fig 9 2 When disinfectant has stopped draining into Bag B cl...

Страница 12: ...ral Information 9 1 Specifications Dimensions Height 11 280 mm width 5 5 140 mm depth 4 5 114 mm excluding IV pole clamp Weight Less than 6 lbs 2 7 kg without water reservoir Vapotherm spike set Works...

Страница 13: ...Appendix A WARNINGS These disinfectant solutions are designed to be used to disinfect the Vapotherm 2000i machine without the cartridge in place These disinfectant solutions ARE NOT approved to disinf...

Страница 14: ...syringe or pipette graduated in 1ml increments Concentrate Disinfectant Minncare Cold Sterilant EPA Reg No 52252 4 22 00 Hydrogen Peroxide 4 50 Peroxyacetic Acid 73 50 Inert Ingredients Control III Di...

Страница 15: ...Vapotherm Inc 198 Log Canoe Circle Stevensville MD 21666 T 001 410 604 3977 F 001 410 604 3978 www vtherm com PN 8 300068 00 Rev E bdc 8369 Supplied by...

Отзывы: