9
If the cycle is interrupted, RF energy is cut off, but the sealing head remains
closed for the duration of the post delay time.
WARNING:
Allow the material to cool before proceeding.
3.1.6.
Once the cycle ends, the generator returns to an idle state.
NOTE:
When spacing seals, make sure that the distance between seals is
at least 10 times
the diameter of the tubing
. Seals that are too close together can cause segments
to rupture.
3.2.
Accessing the Internal Controller
3.2.1.
Ensure that the generator is in an idle state. The LCD screen says “Ready.”
See Figure 2.6.
3.2.2.
Press the green button and red button at the same time until the LCD screen
changes to the password entry screen. See Figure 3.3.
Figure 3.3 LCD password entry screen
The password is a 4-digit numeric code. One digit is active at a time; the
remaining three are masked with * characters.
3.2.3.
Press the green button to increment the active digit. Press the red button to
decrement the active digit.
3.2.4.
Press and hold the green button to activate the next digit (to the right). Press
and hold the red button to activate the previous digit (to the left).
3.2.5.
Press and hold the green button and the red button at the same time to submit
the password.
If the entered password is correct, the LCD screen shows the system
configuration screen.
If the entered password is incorrect, the generator returns to an idle state.
NOTE:
The default password is 0000. To change the password, refer to section 3.5.2.
3.3.
Programming the Internal Controller
3.3.1.
After a valid password is entered, the LCD screen shows the system
configuration screen. See Figure 3.4.