Figure 5:  Adjusting Z height using a Spacer 

 

3.

 

To hold Microplate Platform (13) in 

this “up” position

, tighten Z Arm Nylon Pressure 

Screw (9). 

4.

 

Tighten the Z Bar Stop Thumb Screws (7) to lock the maximum microplate position in 
place. 

5.

 

Loosen the Nylon Pressure Screw (9), lower the Microplate Platform and remove the 
microplate. 

6.

 

Remove the paper spacer from the Microplate Platform and a 1mm space will be set in 
the system. 

Connecting the Manifold to Vacuum System 
1.

 

Connect the vacuum source to a vacuum trap (Figure 6).  
 

 

 

Figure 6:  Vacuum System Connection 

14 

Содержание VP 179A

Страница 1: ...FOR VP 179A MANIFOLD JIG 1 Jig Slot for Shoulder Bolt 2 Shoulder Bolt Head 3 Manifold Hold Down Screw 4 Manifold Hold Down Bar 5 X Positioning Screw 6 Y Position Screw 7 Z Bar Stop Thumb Screw 8 Z Ar...

Страница 2: ...osition always loosen the X Positioning Screws first in the direction the manifold is to be moved before tightening the opposing X Positioning Screws Failure to do this can pull the Shoulder Bolt 2 ou...

Страница 3: ...Stops b Adjusting Nylon Pressure Screw for Z Arm control 2 Slide an empty microplate into the Microplate Platform 13 To set the plate height where the pins stop 1mm from the bottom of the wells place...

Страница 4: ...Stop Thumb Screws 7 to lock the maximum microplate position in place 5 Loosen the Nylon Pressure Screw 9 lower the Microplate Platform and remove the microplate 6 Remove the paper spacer from the Micr...

Страница 5: ...Valve 4 Connect the in line shut off valve to the Vacuum Outlet Fitting 10 with Vacuum Tubing 12 5 Place the Vacuum Tubing 12 in the Support Arm 11 secure to the Jig IMPORTANT NOTE The Aspiration Disp...

Страница 6: ...ATION MANIFOLDS 1 If a single well not emptying properly locate the clogged tube corresponding to the well Insert the rapier metal wire included with the Manifold into the clogged tube This should rem...

Страница 7: ...P Scientific In the event this product fails under normal laboratory conditions within the specified period of time because of a defect in material or workmanship V P Scientific will at its option re...

Отзывы: