V&P Scientific, Inc.  

Technical Note 115A 

SETUP FOR ASPIRATION MANIFOLD ON JIG 

Attaching the Manifold 

1.

 

Loosen Manifold Hold-Down Screws (3) using the 1/16 Allen wrench and remove the 
Manifold Hold Down Bars (4) (Figure 1). 

2.

 

Place the shoulder bolt heads (2) of the Manifold into the Jig slot (1) and slide down 
into the slots until shoulder bolts hit the bottom of the slots. 

3.

 

Place the Manifold Hold-Down Bars (4) on top of the Manifold and tighten the 
Manifold Hold-Down Screws (3) to secure the Manifold in the Jig (Figure 2a). 

Setting the X and Y Positions 

1.

 

Slide a microplate into the Microplate Platform (13).  Use the Allen wrench provided to 
adjust the X and Y Manifold position so it is in the center of the wells of the microplate 
(Figure 2a and 2b).   

2.

 

When adjusting the X Position always loosen the X Positioning Screws first in the 
direction the manifold is to be moved before tightening the opposing X Positioning 
Screws.  Failure to do this can pull the Shoulder Bolt (2) out of the Manifold and strip 
the threads. 

  

 

Figure 2:  Securing Manifold in Jig (a).  Adjusting X (a) and Y (b) Positions. 

 

3.

 

For more precise positioning use one of the Registration Plates VP 903R-96, VP 903R-
384 or VP 903R-1536 (Figure 3).  Place the Registration Plate into the Microplate 
Platform (13) where the microplate would go and push against the back wall.  Pull the 
Z Arm Lever (8) down until the Manifold tubes are just over the Registration Plate.  
Tighten the Z Arm Nylon Press Screw (9) to hold the Microplate Platform (13) in 
position (Figure 4).  Line up the white 1mm corner spots of the Registration Plate with 
the corner tubes on the Manifold.  This is done by using the provided 0.5 Allen wrench 
to adjust the X and Y Position Screws (5 and 6) as shown in Figure 2. 

2a 

2b 

Содержание VP 179A

Страница 1: ...FOR VP 179A MANIFOLD JIG 1 Jig Slot for Shoulder Bolt 2 Shoulder Bolt Head 3 Manifold Hold Down Screw 4 Manifold Hold Down Bar 5 X Positioning Screw 6 Y Position Screw 7 Z Bar Stop Thumb Screw 8 Z Ar...

Страница 2: ...osition always loosen the X Positioning Screws first in the direction the manifold is to be moved before tightening the opposing X Positioning Screws Failure to do this can pull the Shoulder Bolt 2 ou...

Страница 3: ...Stops b Adjusting Nylon Pressure Screw for Z Arm control 2 Slide an empty microplate into the Microplate Platform 13 To set the plate height where the pins stop 1mm from the bottom of the wells place...

Страница 4: ...Stop Thumb Screws 7 to lock the maximum microplate position in place 5 Loosen the Nylon Pressure Screw 9 lower the Microplate Platform and remove the microplate 6 Remove the paper spacer from the Micr...

Страница 5: ...Valve 4 Connect the in line shut off valve to the Vacuum Outlet Fitting 10 with Vacuum Tubing 12 5 Place the Vacuum Tubing 12 in the Support Arm 11 secure to the Jig IMPORTANT NOTE The Aspiration Disp...

Страница 6: ...ATION MANIFOLDS 1 If a single well not emptying properly locate the clogged tube corresponding to the well Insert the rapier metal wire included with the Manifold into the clogged tube This should rem...

Страница 7: ...P Scientific In the event this product fails under normal laboratory conditions within the specified period of time because of a defect in material or workmanship V P Scientific will at its option re...

Отзывы: