© July 2017 Van Essen Instruments. All rights reserved
.
9
3.10
Reg. 00011: Start/Stop Logging of Diver
Setting register 00011 to 0xFF00 will start logging of the Diver and setting this register to 0x0000 will
stop logging of the Diver.
Example:
Description
Function
code
Register
Size
(registers)
Data
field
Remark
1
Set the real-time clock of
the Diver
0x03
40103
6
YY,MM,D
D,
HH,MM,
SS
First register 40103 is YEAR
(YY)
2
Set the sample interval of
the Diver
0x03
40101
2
tt,xx
FI: Register 40101 =01(tt) and
Register 40102=30(xx) gives
30 second sample interval
3
Set the monitoring point
name of the Diver
0x03
40109
10
20 chars Register 40109 - 40118
4
Start the Diver
0x05
00011
1
0xFF00
Register 11: 0xFF00 is start
3.11
Reg. 30241: Number of Records Recorded by Diver
This register contains the number of records recorded by the Diver. First set register 00001 to update
this register in the Diver-MOD with the most recent value.
Example:
Description
Function
code
Register
Size
(registers)
Data
field
Remark
1
Update settings from
Diver in Diver-MOD
0x05
00001
1
0xFF00
2
Read the number of
records recorded by the
Diver
0x04
30241
1
-
Register 30241, integer value
from 0 to 48,000
3.12
Reg. 30242: Maximum Number of Records for Diver
This register contains the maximum number of records that can be recorded by the connected Diver.
Example:
Description
Function
code
Register
Size
(registers)
Data
field
Remark
1
Read the Diver’s memory
capacity in records
0x04
30242
1
-
Register 30242, integer value
24,000 or 48,000 records
3.13
Reg. 30243: Remaining Battery Capacity Diver
This register returns the remaining battery capacity of the connected Diver as a percentage of its full
capacity. This value is only an indication! If the remaining battery capacity comes close to zero, the