© July 2017 Van Essen Instruments. All rights reserved
.
7
Example:
Description
Function
code
Register
Size
(registers)
Data
field
Remark
1
Change baudrate
0x06
40002
1
<1-8>
Register 40002, will take
effect after reboot (reset).
3.5
Reg. 40003: Change Parity of the RS485 port
Register 40003 contains the parity of the RS485 port. The default is no parity. The data field must
have a value from 0 to 3, which corresponds to the following parity:
0 = none
1 = even
2 = odd
3 = mark
Example:
Description
Function
code
Register
Size
(registers)
Data
field
Remark
1
Change parity
0x06
40003
1
<0-3>
Register 40003, will take
effect after reboot (reset).
3.6
Reg. 00001: Update Real-Time Data from Diver
Setting this register forces the Diver-MOD to get real-time values, i.e. pressure, temperature and
conductivity (if applicable), from the connected Diver. The updated real-time values must be read
from registers 30201 – 30216.
The type of conductivity is either normal conductivity or specific conductivity at 25 °C depending on
the setting in the CTD-Diver.
Example:
Description
Function
code
Register
Size
(registers)
Data
field
Remark
1
Update real-time values
from Diver, i.e. copy
values to Diver-MOD
0x05
00001
1
0xFF00
Register 00001 can also be a
broadcast (address 0) to all
Diver-MODs. In case of a
broadcast, wait at least
500 ms before the next
instruction.
2
Read real-time values
from all channels (longs)
0x04
30201
6
-
Register 3020x, convert to
longs (two registers per long)
3
Read real-time values
from all channels (float)
0x04
30211
6
-
Register 3021x, convert to
single floats (two registers
per single float)
4
Go to step 1
When used in a loop