![Van Der Stahl MD-950 NanoPak Series Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/van-der-stahl/md-950-nanopak-series/md-950-nanopak-series_operating-instructions-manual_3342946017.webp)
ENGLISH
MD 950
/ NanoPak
Commissioning
Chapter 2
Page GB 17
9.694.038
V 4.02
2.2 The sealing seam test
Testing of the critical process parameters temperature, contact pressure and sealing time by
means of the sealing seam test. This test should be performed and documented before and
after the daily work process and/or before/after each lot (EN ISO 11607-2).
The sealing parameters are displayed during the tests.
Activating and starting the sealing seam test
1. If a user has been created in the user list, an user
code must be entered before a test is carried out.
Otherwise this is not necessary.
2. Activate the sealing seam test
3. Select the type of sealing seam test
3.1 Seal check
Cannot be used for packages with a side fold!
Place the indicator strips inside the package.
3.2 Ink test
Place the disposable cartridge inside the package
3.3 Peel test
Cut out the test strips after sealing
4. Insert the sterilisation packaging, width min. 200 mm.
This key is used to cancel the function