Testing the Smoke Evacuator
4-6
RapidVac™ Smoke Evacuator Service Manual
10.
Activate the generator in cut mode until the smoke evacuator motor starts.
Depending on the generator, this may take up to 40 seconds.
11.
When the smoke evacuator motor starts, deactivate the generator in cut mode.
12.
Set the generator to coag mode.
13.
Activate the generator in coag mode until the smoke evacuator motor starts.
Depending on the generator, this may take up to 40 seconds.
14.
When the smoke evacuator motor starts, deactivate the generator in coag
mode.
The Remote Switch Activator is now programmed. Powering on and off the
RapidVac Smoke Evacuator is now controlled by the generator and Remote
Switch Activator.
Installing and Testing the Generator Interlink Cable (optional)
The Generator Interlink Cable attaches to a RapidVac Smoke Evacuator and a
compatible Valleylab generator to synchronize smoke evacuation with the
activation of the generator. The cable is compatible with the Valleylab generator
models Force FX™, ForceEZ™, and ForceTriad™.
1.
Turn power off and unplug the smoke evacuator and generator from the power
source.
2.
Remove the metal cover plate in the center of the generator back panel to
access optional ports.
3.
Plug the interlink cable into the generator RF Activation port.
4.
Plug the interlink cable into the Remote Activator jack on the back panel of
the smoke evacuator.
Important
Do not activate the generator in cut mode again until the Remote Switch Activator
has been programmed in coag mode.
Important
If the generator has only one tone, program this same tone into the Remote
Switch Activator memory.
If both the cut and coag tones are not activating the smoke evacuator reprogram
the Remote Switch Activator. Disconnect the Remote Switch Activator from the
smoke evacuator, then re-connect. Repeat the programming instructions
beginning with step 8.
Important
If the attached smoke evacuator is to be disconnected from the generator, unplug
the Remote Switch Activator from the smoke evacuator. The Remote Switch
Activator should remain affixed to the generator to which it was programmed. The
RapidVac Smoke Evacuator cannot be operated manually or with a footswitch
while the Remote Switch Activator is plugged into the smoke evacuator.
If the Remote Switch Activator is moved to another generator, it must be
reprogrammed using the steps in this section.
Содержание RapidVac 1009803
Страница 1: ...Service Manual RapidVac Smoke Evacuator...
Страница 6: ...vi RapidVac Smoke Evacuator Service Manual...
Страница 36: ...4 10 RapidVac Smoke Evacuator Service Manual...
Страница 40: ...5 4 RapidVac Smoke Evacuator Service Manual...
Страница 44: ...6 4 RapidVac Smoke Evacuator Service Manual...