EMC for the RapidVac Smoke Evacuator
T
e
chnical S
pecifications
RapidVac™ Smoke Evacuator Service Manual
3-3
Power
This unit is equipped from the factory with either a 110 VAC hospital grade
NEMA 5-15 power cord or a 220 VAC CEE7/7 power cord. Should the AC
power cord need replacing to match another plug configuration, the replacement
plug/cable/receptacle configuration must meet or exceed the following
specifications.
100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft. (5m)
Plug - minimum 10A 125 VAC
Unit receptacle - IEC female, minimum 10A 125 VAC
220-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15 ft. (5m)
Plug - minimum 6A 250 VAC
Unit receptacle - IEC female, minimum 6A 250 VAC
Operating Parameters
Transport and Storage
EMC for the RapidVac Smoke Evacuator
Electromagnetic Compatibility (IEC 60601-1-2)
The RapidVac Smoke Evacuator complies with the appropriate IEC 60601-1-2
specifications regarding electromagnetic compatibility.
The RapidVac Smoke Evacuator meets the following requirements:
•
ESD Immunity (IEC 60601-1-2 sub-clause 36.202 and IEC 61000-4-2)
•
Radiated Immunity (IEC 60601-1-2 sub-clause 36.202.2 and
IEC 61000-4-3)
•
Electrical Fast Transient/Burst (IEC 60601-1-2 sub-clause 36.202.3.1 and
IEC 61000-4-4)
•
Surge Immunity (IEC 60601-1-2 sub-clause 36.202.3.2 and
IEC 61000-4-5)
•
Emissions (IEC 60601-1-2 sub-clause 36.201.1)
Ambient temperature range
10°C to 40°C (50°F to 104°F)
Relative humidity
30% to 75%, noncondensing
Ambient temperature range
-34°C to 65°C (-29°F to 149°F)
Relative humidity
25% to 85%, noncondensing
Содержание RapidVac 1009803
Страница 1: ...Service Manual RapidVac Smoke Evacuator...
Страница 6: ...vi RapidVac Smoke Evacuator Service Manual...
Страница 36: ...4 10 RapidVac Smoke Evacuator Service Manual...
Страница 40: ...5 4 RapidVac Smoke Evacuator Service Manual...
Страница 44: ...6 4 RapidVac Smoke Evacuator Service Manual...