Testing the Smoke Evacuator
Setup, T
e
st
s, and Maintenance
RapidVac™ Smoke Evacuator Service Manual
4-5
3.
Turn on the smoke evacuator by pressing the power on switch ( | ) on the
lower left of the front panel. The smoke evacuator powers up with the fan
motor off (standby mode). The LED in the center of the fan symbol blinks
indicating power is on.
4.
Press the up-arrow (
) button once to start suction at the lowest air-flow
setting.
5.
Press the up-arrow (
) button and pause briefly to progress to the next level
of suction. Verify that the level of suction noticeably increases with each
progression.
6.
When all suction flow LEDs are lit steadily (100% motor speed), press the
down-arrow (
) button repeatedly to regress through the settings to the
lowest setting.
Installing and Testing the Footswitch (optional)
Only use the Valleylab Footswitch with the smoke evacuator. Other footswitches
are not compatible.
1.
Turn on the smoke evacuator.
2.
Connect the footswitch cord to the footswitch jack on the front panel.
3.
Turn on the airflow to any setting.
4.
Press the footswitch once to turn off the airflow.
5.
Press the footswitch again to activate the airflow.
Installing and Testing the Remote Switch Activator (optional)
Use the Valleylab Remote Switch Activator with any of Valleylab’s
electrosurgical pencils and generators.
1.
Select a location on the generator near the speaker where the Remote Switch
Activator is to be mounted.
2.
Clean the surface of the selected area with alcohol. Allow the alcohol to
evaporate.
3.
Remove the backing from the gasket on the RSA.
4.
Affix the RSA to the generator.
5.
Plug the RSA connector into the Remote Activator jack located on the back
panel of the smoke evacuator.
6.
Turn on the smoke evacuator and press the up-arrow (
) button twice to
illuminate the first two air flow indicators (40% motor speed).
7.
Set up the surgical pencil and generator as instructed in their user
documentation. If the generator’s audible-tone volume can be adjusted, set it
at 50%.
8.
Initiate programming the Remote Switch Activator by pressing the
Program/Reset button continually for 10 seconds.
9.
Set the generator to cut mode.
Содержание RapidVac 1009803
Страница 1: ...Service Manual RapidVac Smoke Evacuator...
Страница 6: ...vi RapidVac Smoke Evacuator Service Manual...
Страница 36: ...4 10 RapidVac Smoke Evacuator Service Manual...
Страница 40: ...5 4 RapidVac Smoke Evacuator Service Manual...
Страница 44: ...6 4 RapidVac Smoke Evacuator Service Manual...