RESISTANCE ADJUSTMENTS 

1.

 

Select 

200Ω

 

range. 

2.

 

Connect to a 

100Ω 

standard resistor. 

Adjust R15 for a display reading equal to the value of the resistor. 

3.

 

Select 

2kΩ r

ange and connect to 

a 1kΩ

 standard resistor. 

Adjust RV1 for a display reading equal to the resistor value. 

4.

 

Select 

20kΩ range

 and connect to 

a 10kΩ

 standard resistor. 

Adjust RV2 for a display reading equal to the resistor value. 

5.

 

Select 200kΩ range and connect to a 100kΩ standard resistor.

 

Adjust RV3 for a display reading equal to the resistor value. 

6.

 

Select 2MΩ range and connect to a 1MΩ standard resistor.

 

No adjust. Verify that display reading is within tolerance of the resistor value. 

7.

 

Select 2Ω range and connect to a 1Ω standard resistor.

 

No adjust. Verify that display reading is within tolerance of the resistor value. 

8.

 

If the 2Ω range is too low select R27 using the RC Box.

 

9.

 

Select 20Ω range and connect to a 10Ω standard resistor.

 

No adjust. Verify that display reading is within tolerance of the resistor value. 

10.

 

If the 20Ω range is too low select R

37 using the RC Box. 

 

 
IMPORTANT:

 If it is necessary to perform step 4.8 or 4.10, go back and forth between the 

2Ω, 20Ω and 

200Ω range

s until all are in spec. 

 

FAILSAFE CHECK 

 

1.

 

Connect the 3458A in the ammeter mode across the 
current terminals. 

Select the 2Ω range

2.

 

Short out the cathode of D9 and the bottom pin or R44 
(see image)  and  verify that the measured current is 
equal or less then the value specified in the Fail Safe 
current table. 

3.

 

Repeat this process for each range. 

4.

 

For  MΩ  ranges  it  will  be  necessary  to  connect  the 

appropriate calibration resistor and  measure the 
voltage drop across it. Use Ohm’s Law to calculate the 
failsafe current. 
 

Range 

Fail-Safe Current 

2Ω

 

8mA 

20Ω

 

8mA 

200Ω

 

1.8mA 

2kΩ

 

180µA 

20kΩ

 

18µA 

200kΩ

 

1.8µA 

2MΩ

 

180nA 

 

 

Figure 

7 – Fail-Safe shortin

g points

 

Содержание 4314KV

Страница 1: ...4314KV 4314KVS Rugged Digital Igniter Tester Operations Manual Valhalla Scientific Inc 12127 Kirkham Rd Poway CA 92064 www valhallascientific com...

Страница 2: ......

Страница 3: ...this instrument is stated on your invoice and packing list Please refer to these to determine appropriate warranty dates We will repair the instrument during the warranty period provided it is returne...

Страница 4: ...al shock DO NOT perform procedures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features buil...

Страница 5: ...cations 4 Chapter 4 Optional Equipment 5 Available Options 5 Option A2 Battery Charger 5 Replacement Batteries 5 AC 115 Continuous Isolated AC Power 5 Test Leads 5 Option K 174 4 Wire Kelvin Lead Set...

Страница 6: ...11 Interpreting the Display 11 Current Display and Multiplier LEDs KVS models only 11 Fail Safe Operation 12 Fail Safe Test model 4314KVS only 12 Low Battery Indicator 13 Standby Mode 13 Chapter 6 Tr...

Страница 7: ...19 Chapter 8 Structural Features 20 Certifications 20 Chapter 9 Addendums 21 Table of Figures Figure 1 Fail Safe test button detail KVS 7 Figure 2 4314KV Front Panel 8 Figure 3 4314KVS Front Panel 9 F...

Страница 8: ...designed in a compact rugged water resistant Pelican case for protection in the harshest environments Measures Under all Conditions The 4314KV KVS are not only the safest resistance testers for explos...

Страница 9: ...eavy duty Nickel metal hydride NiMH battery pack The battery charge is maintained by an external AC DC converter that plugs into a standard receptacle The AC adapter can accept an input voltage from 1...

Страница 10: ...ailsafe Currents vary from instrument to instrument The actual measured level is provided on the certificate of calibration Accuracy Table 2 Range Accuracy 2 0 03 of reading 0 02 of range 20 0 03 of r...

Страница 11: ...nt Display KVS model 3 digits Light Emitting Diodes LED 999 Overload Indication Screen displays O L Conversion Rate 3 readings per second Terminal Configuration 4 Wire Kelvin Current Source Compliance...

Страница 12: ...ircuits When this option is installed the 4314KV KVS is no longer battery powered 100 240Vac 50 60Hz Test Leads Option K 174 4 Wire Kelvin Lead Set RG174 coaxial cable Option K 174 comes standard with...

Страница 13: ...via a spring loaded center contact The MP 4 target area is 1 in diameter The MP 5 target area is 4 in diameter Option KK 4 Wire Kelvin Lead Set Option KK is a 48 heavy duty cable set terminated in lar...

Страница 14: ...d to place the tester in Stand By mode Fuse Holder The fuse holder is mounted on the front panel and contains a 2 amp in line fuse This fuse is designed to protect the battery pack from excessive char...

Страница 15: ...8 Figure 2 4314KV Front Panel...

Страница 16: ...9 Figure 3 4314KVS Front Panel...

Страница 17: ...potential error source by providing two terminals of constant current and an additional two terminals for high impedance voltage measurement The result is a fast accurate resistance measurement of the...

Страница 18: ...ring display is not acceptable the ohmmeter can be made to indicate zero whenever the sense terminals are shorted The display should indicate a stable reading when the test leads are securely attached...

Страница 19: ...orst case failure condition The resulting output current is recorded and provided on the Certificate of Calibration included with every unit Fail Safe Test model 4314KVS only The 4314KVS is equipped w...

Страница 20: ...usted An overnight recharge should be performed before using the device for critical testing NOTE It is possible for the user to receive a low battery indication on a single range only particularly th...

Страница 21: ...or on a specific range only The power supplies are also one of the first things that should be checked Schematics and Parts List The schematics and parts list are not part of this User Manual Please c...

Страница 22: ...15 Block Diagram Figure 6 Block Diagram...

Страница 23: ...st Equipment Following is a list of the standard resistors and other equipment required to calibrate the 4314KV KVS Precision Resistors 1 0 01 Accuracy 1 0 005 Accuracy 10 0 005 Accuracy 100 0 005 Acc...

Страница 24: ...V and press the FAILSAFE TEST button 3 When the display prompts CAL 3 Apply 75mV and press the FAILSAFE TEST button 4 When the display prompts CAL C 5 00mA Remove voltage connections and connect the S...

Страница 25: ...the 2 range is too low select R27 using the RC Box 9 Select 20 range and connect to a 10 standard resistor No adjust Verify that display reading is within tolerance of the resistor value 10 If the 20...

Страница 26: ...and install batteries Checking the Calibration Results Scale and Linearity Check Apply to each range resister values equal to 50 of range 5 of range 0 5 of range and 0 05 of range All values should b...

Страница 27: ...alve balances interior pressure keeps water out Easy open Double Throw latches Certifications Certified Mil Std 810 Method 511 Safe for Explosive Volatile Atmospheres Complies to EWR 127 1 Section 3 1...

Страница 28: ...ating considerations for special 4314KV KVS Please refer to any applicable material before attempting to operate your Model 4314KV KVS If you have a digital copy of this manual all available addendums...

Страница 29: ...lascientific com For information on our line of products please visit or web site at www valhallascientific com To reach our sales department Phone 858 457 5576 Fax 858 457 0127 E mail sales valhallas...

Страница 30: ...23...

Страница 31: ......

Страница 32: ...4314KV and 4314KVS Digital Igniter Tester User Manual Revision 1 Edition 1 2017 Copyright 2017 Valhalla Scientific Inc All rights reserved...

Отзывы: