Chapter 5 – Features and Operations 

General 

This section of the manual contains complete operating instructions for the Model 4314KV/KVS Digital 
Igniter Testers. A description of the front panel controls, connection instructions, and the theory behind 
4-wire resistance measurement is discussed in this section. 

 

Front Panel 

ON/OFF Button 

When the front panel power switch is placed in the OFF/CHARGE position, all power is removed from the 
output terminals, and the battery pack is connected to the charging circuit. When the switch is placed in 
the ON position, the battery pack is disconnected from the charging circuit. The possibility of a common 
mode voltage between the device under test and AC Power ground is therefore eliminated. 

 

Range Selection Knob 

The 4314KV/KVS measurement ranges are selected by turning the range knob to the desired position. 
Move knob clockwise to increase the range and counterclockwise to reduce the range. Range knob is also 
used to place the tester in Stand By mode. 

 

Fuse Holder 

The fuse holder is mounted on the front panel and contains a 2 amp in-line fuse. This fuse is designed to 
protect the battery pack from excessive charging currents. For continued protection, replace the line fuse 
only with fuse of the same voltage and current rating and type. DO NOT use  repaired fuses or short-
circuited fuse holders. 

 

Charging Jack 

The battery charging jack is a barrel type and is located on the 4314KV/KVS front panel. The center pin of 
the connector is positive. The charging requirements of the internal battery pack are 6VDC @ 1.25A. The 
correct charging voltage is supplied by the adapter included with the instrument. Additional adapters are 
available as Option "A2". Charger is equipped with a locking ring, and is sealed IP68.  
 

Fail Safe Button (model 4314KVS only) 

When pressed, the Fail-Safe test button simulates a worst case component failure and displays the “real” 
fail-safe current. During the test, the current source is isolated from the binding-post.  The ohms 
measurement window will display FSt, to indicate that the unit is in fail-safe test mode. 
 

 

Figure 

1 - Fail-Safe test button detail (KVS) 

Содержание 4314KV

Страница 1: ...4314KV 4314KVS Rugged Digital Igniter Tester Operations Manual Valhalla Scientific Inc 12127 Kirkham Rd Poway CA 92064 www valhallascientific com...

Страница 2: ......

Страница 3: ...this instrument is stated on your invoice and packing list Please refer to these to determine appropriate warranty dates We will repair the instrument during the warranty period provided it is returne...

Страница 4: ...al shock DO NOT perform procedures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features buil...

Страница 5: ...cations 4 Chapter 4 Optional Equipment 5 Available Options 5 Option A2 Battery Charger 5 Replacement Batteries 5 AC 115 Continuous Isolated AC Power 5 Test Leads 5 Option K 174 4 Wire Kelvin Lead Set...

Страница 6: ...11 Interpreting the Display 11 Current Display and Multiplier LEDs KVS models only 11 Fail Safe Operation 12 Fail Safe Test model 4314KVS only 12 Low Battery Indicator 13 Standby Mode 13 Chapter 6 Tr...

Страница 7: ...19 Chapter 8 Structural Features 20 Certifications 20 Chapter 9 Addendums 21 Table of Figures Figure 1 Fail Safe test button detail KVS 7 Figure 2 4314KV Front Panel 8 Figure 3 4314KVS Front Panel 9 F...

Страница 8: ...designed in a compact rugged water resistant Pelican case for protection in the harshest environments Measures Under all Conditions The 4314KV KVS are not only the safest resistance testers for explos...

Страница 9: ...eavy duty Nickel metal hydride NiMH battery pack The battery charge is maintained by an external AC DC converter that plugs into a standard receptacle The AC adapter can accept an input voltage from 1...

Страница 10: ...ailsafe Currents vary from instrument to instrument The actual measured level is provided on the certificate of calibration Accuracy Table 2 Range Accuracy 2 0 03 of reading 0 02 of range 20 0 03 of r...

Страница 11: ...nt Display KVS model 3 digits Light Emitting Diodes LED 999 Overload Indication Screen displays O L Conversion Rate 3 readings per second Terminal Configuration 4 Wire Kelvin Current Source Compliance...

Страница 12: ...ircuits When this option is installed the 4314KV KVS is no longer battery powered 100 240Vac 50 60Hz Test Leads Option K 174 4 Wire Kelvin Lead Set RG174 coaxial cable Option K 174 comes standard with...

Страница 13: ...via a spring loaded center contact The MP 4 target area is 1 in diameter The MP 5 target area is 4 in diameter Option KK 4 Wire Kelvin Lead Set Option KK is a 48 heavy duty cable set terminated in lar...

Страница 14: ...d to place the tester in Stand By mode Fuse Holder The fuse holder is mounted on the front panel and contains a 2 amp in line fuse This fuse is designed to protect the battery pack from excessive char...

Страница 15: ...8 Figure 2 4314KV Front Panel...

Страница 16: ...9 Figure 3 4314KVS Front Panel...

Страница 17: ...potential error source by providing two terminals of constant current and an additional two terminals for high impedance voltage measurement The result is a fast accurate resistance measurement of the...

Страница 18: ...ring display is not acceptable the ohmmeter can be made to indicate zero whenever the sense terminals are shorted The display should indicate a stable reading when the test leads are securely attached...

Страница 19: ...orst case failure condition The resulting output current is recorded and provided on the Certificate of Calibration included with every unit Fail Safe Test model 4314KVS only The 4314KVS is equipped w...

Страница 20: ...usted An overnight recharge should be performed before using the device for critical testing NOTE It is possible for the user to receive a low battery indication on a single range only particularly th...

Страница 21: ...or on a specific range only The power supplies are also one of the first things that should be checked Schematics and Parts List The schematics and parts list are not part of this User Manual Please c...

Страница 22: ...15 Block Diagram Figure 6 Block Diagram...

Страница 23: ...st Equipment Following is a list of the standard resistors and other equipment required to calibrate the 4314KV KVS Precision Resistors 1 0 01 Accuracy 1 0 005 Accuracy 10 0 005 Accuracy 100 0 005 Acc...

Страница 24: ...V and press the FAILSAFE TEST button 3 When the display prompts CAL 3 Apply 75mV and press the FAILSAFE TEST button 4 When the display prompts CAL C 5 00mA Remove voltage connections and connect the S...

Страница 25: ...the 2 range is too low select R27 using the RC Box 9 Select 20 range and connect to a 10 standard resistor No adjust Verify that display reading is within tolerance of the resistor value 10 If the 20...

Страница 26: ...and install batteries Checking the Calibration Results Scale and Linearity Check Apply to each range resister values equal to 50 of range 5 of range 0 5 of range and 0 05 of range All values should b...

Страница 27: ...alve balances interior pressure keeps water out Easy open Double Throw latches Certifications Certified Mil Std 810 Method 511 Safe for Explosive Volatile Atmospheres Complies to EWR 127 1 Section 3 1...

Страница 28: ...ating considerations for special 4314KV KVS Please refer to any applicable material before attempting to operate your Model 4314KV KVS If you have a digital copy of this manual all available addendums...

Страница 29: ...lascientific com For information on our line of products please visit or web site at www valhallascientific com To reach our sales department Phone 858 457 5576 Fax 858 457 0127 E mail sales valhallas...

Страница 30: ...23...

Страница 31: ......

Страница 32: ...4314KV and 4314KVS Digital Igniter Tester User Manual Revision 1 Edition 1 2017 Copyright 2017 Valhalla Scientific Inc All rights reserved...

Отзывы: