Pin-type Test Lead Zero-Adjustment

4. 

When the resistance reading is lower than 200 digits and stable, the zero-adjustment is automatically carried out. The 

            would then stop flashing, and the beeper sound would also stop. 

5. 

Remove the pin-type test leads from the zero-adjust board and start the measurement, connect the leads to the 

battery to test.

6. 

Press the          key to exit this function.  

NOTE:

 if the terminal spacing of the battery to be measured is larger than the zero-adjustment board, use the outer 

most holes for adjustment. The zero-adjustment s valid for the currently selected range only, as long as the power 

remains on. Powering on the meter would reset all zero-adjustment values. 

Clip-type Test Lead Zero-Adjustment

Short the SOURCE and SENSE of the clip-type test lead with the temperature sensor together as shown in the image: 

SOURCE

SENSE

Hold and Auto-Hold Functions

Hold Function

Press the          key,            appears on display is locked. Press the          key again to exit. 

Auto-Hold Function

1. 

Press the           key for 2 seconds to start the auto-hold function,                    appears on the display. 

2. 

The meter holds the reading on the display until it detects a new stable reading. The meter beeps and displays the new 

reading. When the measured reading is stable, 

 appears on display. 

3. 

Press the           key again to exit. 

Содержание RE856

Страница 1: ...AUSTRALIA 1300 734 253 sales valen com au www valen com au NEW ZEALAND 0800 734 253 sales valen co nz www valen co nz OPERATION MANUAL RE856 BATTERY TESTER...

Страница 2: ...ry increasing growth the Lake Cargelligo Head Office moved into a brand new purpose built facility in 2012 In 2015 Regal Electro successfully completed its rst ENLiFEN Energy Storage Optimiser ESO pro...

Страница 3: ...done Upbeat Positive Attitude If our employees have a positive attitude and constantly strive to give your best effort eventually we will overcome our immediate problems and nd you are ready for grea...

Страница 4: ...nals 6 Screen Display 7 SPECIFICATIONS 8 General Specifications 8 Electrical Specifications 9 OPERATION 10 Preparation 10 Operation 10 Operation 11 Zero adjust Function 11 Pin type Test Lead Zero Adju...

Страница 5: ...RE856 OPERATION MANUAL 5 AUS 1300 734 253 NZ 0800 734 253 NAMES FUNCTIONS OF PARTS 1 2 3 6 9 4 5 7 8 10 17 11 13 14 12 15 16...

Страница 6: ...emory and auto data logging interval time setting mode press this key again to exit 7 key To turn the comparator function on and off Hold this key for 2 seconds to set the comparator setting mode pres...

Страница 7: ...omparator voltage low limit setting Measured battery charge capacity Measured battery discharge capacity Measured battery total capacity The calculated remaining capacity of the battery Shown when the...

Страница 8: ...isplay of or results and beeper for warning and fail results Low In High Low High Voltage Resistance Manual and auto data memory 999 sets can be read by meter and downloaded by PC Auto data logging Mi...

Страница 9: ...easurement Measuring current frequency 1kHz 30Hz Range Resolution Accuracy 6V 1mV 0 1 readings 6 digits 60V 10mV Voltage measurement Maximum input voltage 60VDC maximum no AC voltage input DO NOT EXCE...

Страница 10: ...a high voltage battery before continuing to measure a low voltage battery first short the measurement leads together This will discharge the DC elimination capacitor which is connected across the lead...

Страница 11: ...compared with the measurement range Zero adjust Function The zero adjustment function is to zero the range of resistance The reading during zero adjustments would be taken as zero and will be used to...

Страница 12: ...or adjustment The zero adjustment s valid for the currently selected range only as long as the power remains on Powering on the meter would reset all zero adjustment values Clip type Test Lead Zero Ad...

Страница 13: ...Connect the yellow miniature plug of the DCA current adaptor to the meter DCA CLAMP ADAPTOR jack 2 USE DCA CURRENT ADAPTOR Press the key of the DCA Current Adaptor to turn on the adaptor the POWER LE...

Страница 14: ...be The unit then displays with corresponding symbol and sounds a beeper with the WARNING and FAIL cases Comparator Settings 1 Rotate the resistance range switch to any range 2 Press and hold down the...

Страница 15: ...3 Press and keys to select the desired comparator number 01 99 The selected comparator number remains in memory even when the power is turned off 4 Press the key again to stop the comparator function...

Страница 16: ...y charge discharge remained capacity is Data Memory Function Manual and Auto Data Memory Function 999 sets To clear the Memorised Data When the memory is full the memory number 999 will appear on the...

Страница 17: ...splay will show NO XXX 2 Press the and keys to scroll through the memorised readings 3 Press key again to stop viewing the memorised readings 4 The memorised data can be read by transferring to PC Aut...

Страница 18: ...her battery fluid leakage possibility When the meter will not be in use for an extended period of time please remove the batteries out of the meter to prevent the possibility of battery fluid leakage...

Страница 19: ...User Account Control windows opens asking Do you want to allow the following program from an unknown publisher to make changes to this computer click Yes d e Online Operation 1 Click Run BatCapTester...

Страница 20: ...n A dialogue box will open input the file name select a file and then click the button Download Data Download Data from Hard Disk Click the button seventh from the left on the main screen Download Dat...

Страница 21: ...crosoft Excel and locate the saved file in Excel format for example test xls The Text Import Wizard then appears Follow the steps from 1 to 3 to complete Data Convert Excel Open a saved data file in t...

Страница 22: ...mpling Time PC Sampling Time Meter sampling rate is the rate at which the meter stores readings Click the button fifth from the left on the main menu Meter Sampling Rate Click the sixth from left on t...

Страница 23: ...om the right on the main menu Click to set the meter Set Comparator To erase the EEPROM click the button second from the right on the main menu Click to confirm Erase EEPROM To delete the SD card memo...

Отзывы: