RE856
OPERATION MANUAL
AUS
1300 734 253 NZ 0800 734 253
12
Pin-type Test Lead Zero-Adjustment
4.
When the resistance reading is lower than 200 digits and stable, the zero-adjustment is automatically carried out. The
would then stop flashing, and the beeper sound would also stop.
5.
Remove the pin-type test leads from the zero-adjust board and start the measurement, connect the leads to the
battery to test.
6.
Press the key to exit this function.
NOTE:
if the terminal spacing of the battery to be measured is larger than the zero-adjustment board, use the outer
most holes for adjustment. The zero-adjustment s valid for the currently selected range only, as long as the power
remains on. Powering on the meter would reset all zero-adjustment values.
Clip-type Test Lead Zero-Adjustment
Short the SOURCE and SENSE of the clip-type test lead with the temperature sensor together as shown in the image:
SOURCE
SENSE
Hold and Auto-Hold Functions
Hold Function
Press the key, appears on display is locked. Press the key again to exit.
Auto-Hold Function
1.
Press the key for 2 seconds to start the auto-hold function, appears on the display.
2.
The meter holds the reading on the display until it detects a new stable reading. The meter beeps and displays the new
reading. When the measured reading is stable,
appears on display.
3.
Press the key again to exit.