User's Guide ______________________________________________________________________
30 _______________________________________________________________ M210701EN-A
Connections and Selecting the Calibrator Function
1.
To select the calibrator function from the HMI41, press
ON/OFF
button until you can see some text on the display. Then release
the
ON/OFF
button and within 1...2 seconds, press both
ENTER
and
MODE
buttons until the following display appears:
2.
After a few seconds, the display changes to show the following:
set
°C
If the basic settings (display units, automatic power off function,
display quantities, and pressure) have to be changed, please refer
to the HMI41 Operating Manual. Otherwise, press
ENTER
repeatedly until the following display appears:
set
To calibrate the HMT100, select number 3 with buttons
▲
(number up) and
▼
(number down) and then press
ENTER
.
3.
Next the baud rate appears on the display. Use the baud rate
19200 (= 19.2 on the HMI41 display) with the HMT100 series.
If the baud rate on the HMI41 display is not correct, change it
with buttons
▲
and
▼
. When the baud rate is correct, press
ENTER
, and the serial communications settings display appears.
The correct settings for HMT100 series are: N, 8, 1. If needed,
change the settings with buttons
▲
and
▼
until they are
correct, press
ENTER
, and then
ON/OFF
.
These setting are stored in the HMI41 memory; when the HMI41 is
turned on again, it will automatically wake up as a calibrator for
digital transmitters with these serial line settings. After making these
settings, continue with the adjustment and follow the directions of the
chosen adjustment method.