![Vaisala HUMICAP HMT100 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/vaisala/humicap-hmt100/humicap-hmt100_user-manual_3334648020.webp)
User's Guide ______________________________________________________________________
18 _______________________________________________________________ M210701EN-A
Probe Assembly with Duct Installation Kit
2
1
3
0505-177
Figure 8
Assembly of the Humidity Probe with the Duct
Installation Kit
The following numbers refer to Figure 8 above:
1
=
HMP100
2
=
Duct installation kit
3
=
Probe cable
1.
Slide the probe cable through the duct installation kit plastic
pipe.
2.
Attach the probe cable to the HMP100 Humidity Probe.
3.
Attach probe assembly to the duct.
Drilling Instructions for Duct Installation Kit
42
1
2
3
Ø24
Ø3.
2
Drill the holes for the duct installation kit as
follows:
1.
Use a 24 mm drill bit to drill a hole to
the duct wall for the humidity probe.
2.
Drill holes for the duct installation kit
mounting screws around the hole in a
square arrangement, 42 mm apart from
each other. Use a 3.2 mm drill bit to
drill the holes for the mounting screws
(four ST4.2×16-C-Z DIN 7981
screws).
0505-178
Figure 9
Drilling Instructions