• The
ERRS
command returns the currently active error code(s).
For example:
>
errs
Error: E2 Humidity sensor open circuit/not calibrated
>
The analog output levels for the error state are configurable using the
AERR
command;
see
Set Analog Output Error Notification (page 35)
.
6.3 Error Codes
The transmitter has error codes that can be displayed on the serial line using the
ERRS
command. See
Display Currently Active Errors (page 37)
.
Table 8 Error Codes
Error Code
Error Message
Possible Cause(s)
E1
Temperature sensor open
circuit/not calibrated
• Temperature sensor damaged
• Temperature sensor short circuit
• Temperature sensor missing
E2
Humidity sensor open circuit/not
calibrated
• Condensation on humidity sensor
• Humidity sensor damaged
• Humidity sensor short circuit
• Humidity sensor missing
E5
Capacitance out of range
• Condensation on humidity sensor
• Humidity sensor damaged
E6
Program checksum error
Internal error
E7
Parameter storage checksum error
Internal error
E9
Sensor coefficient checksum error
Internal error
E10
Current settings checksum error
Internal error
E11
Factory settings checksum error
Internal error
E12
Temperature sensor values out of
range
• Temperature sensor damaged
• Temperature sensor missing
E14
MeasTask not calculating (ATEST)
This error may occur when the analog output is in
test mode.
See
Test Analog Output (page 38)
.
E16
Default settings not found
Internal error
E17
RH reading too high
Internal error
DMT132 User Guide
M211289EN-B
60