For example:
29.9213 ���� = 29.9213 × 0.03386388 = 1.01325 ���
�
Conversions from mmHg and inHg are defined at 0 °C and from mmH
2
O and inH
2
O at
4 °C.
You can also set the pressure compensation value when you connect the DMT132 to the
MI70 indicator. When using the MI70, the pressure unit is bar
a
.
4.7 Serial Line Output Commands
4.7.1 Output a Reading Once
Use the
SEND
command to output a single measurement message. The output message is
sent for the measurement parameter that is currently selected using the
ASEL
command.
SEND
<cr>
Example:
>
send
Tdf= -0.80'C
>
4.8 Troubleshooting and Maintenance
Commands
4.8.1 Display Currently Active Errors
Use the
ERRS
command to display the currently active transmitter error codes. Possible
errors and their causes are listed in
Table 8 (page 60)
.
ERRS
<cr>
Chapter 4 – Operation
37