rhcal
RH: 11.278554 Ref1 ?
11.3
Press any key when ready
OK
Saving settings...done
>
Example (two-point calibration):
>
rhcal
RH: 11.278554 Ref1 ?
11.3
Press any key when ready
RH: 75.091213 Ref2 ?
75
OK
Saving settings...done
>
4.9.2 Calibrate Temperature Measurement
Use the
CT
command to perform a one-point or two-point temperature (T) calibration.
CT
<cr>
When performing a one-point calibration, you need to place the transmitter in a single
temperature reference. Run the command and enter the exact temperature of the reference
after the measurement has stabilized. Exit the command by pressing the ESC key before
proceeding to the second point (see first example below). To update the value, press ENTER
without inputting a value.
For two-point calibration, the second reference point must be at least 20 °C warmer than
the first. The measurement reading for the second point will appear when this requirement
is met.
Example (one-point calibration):
>
ct
T 22.3104 Ref1 ?
22.3
Press any key when ready
OK
Saving settings...done
>
Example (two-point calibration):
DMT132 User Guide
M211289EN-B
40