4. Verify that the connection works by giving the
?
command:
>
?
If the transmitter does not respond with device information:
• Check your serial line settings and cable connections
• Reconnect the USB cable, restart the terminal application, and try again
5. When your serial connection is working, use the
L
command to see the current user
adjustment parameters.
Write down the RH offset and gain values so you can restore them manually using the
LI
command, if necessary.
>
l
6. Use the
ERRS
command to see that no errors are active:
>
errs
7. Install the HMK15 calibrator fitting over the LiCl salt chamber of the humidity calibrator
(11 %RH).
Make sure it sits tightly over the salt chamber.
8. Insert the DMT132 into the chamber. Do not remove the filter.
Make sure all unused holes are plugged and the transmitter is installed tightly on the
thread.
9. Give the
RHCAL
command.
The output from the command shows the currently measured RH value, and prompts
you to enter the value of the first reference (Ref1). This is the reference on the dry end,
which produces the offset correction for the measurement.
>
rhcal
RH: 10.840536 Ref1 ?
10. Wait for 20 ... 40 minutes for the reading to stabilize.
Press ENTER without typing anything to update the value shown by the
RHCAL
command.
RH: 10.890510 Ref1 ?
RH: 10.920197 Ref1 ?
RH: 10.930460 Ref1 ?