Calibration and adjustment____________________________________________________________
Vaisala _________________________________________________________________________ 45
CHAPTER 10
CALIBRATION AND ADJUSTMENT
Calibration interval
HMT360 is calibrated as shipped from the factory. Typical humidity calibration
interval is one year. Depending on the application it may be good to make the first
calibration check earlier.
Factory calibration and adjustment
The device (or only the electronics unit) can be sent to Vaisala Service Centers for
calibration and adjustment, see contact information on page 9.
HMT360 for
natural gas shall be adjusted always in Vaisala Service Center.
Calibration and adjustment by the user
Note that HMT360 for natural gas can be checked but not adjusted by a user.
Calibration is carried out by using the keypad or by using serial commands. The
following additional equipment is needed in calibration of the HMT360:
•
power supply for the channel(s) to be calibrated (12-24 VDC)
•
a multimeter for the HMT360 without a display
•
the calibrated references
Firstly, the electronics unit is removed from the transmitter base to perform a
calibration in safe area. Secondly, the power supply and a multimeter is connected
to the electronics unit according to the following instructions.