Installation_________________________________________________________________________
Vaisala _________________________________________________________________________ 16
F
IGURE
3-3
Installation of the sensor head HMP363 in a channel
with a flange and a supporting bar
The HMP365 probe for high
temperatures
When installing the HMP365 sensor head in a duct or channel, there
should not be remarkable difference between temperatures inside and
outside of the duct. This can cause incorrect humidity readings.
F
IGURE
3-4
Mounting the HMP365 sensor head in a duct or
channel.
HMP364 for pressurized spaces
The HMP364 pressure tight probe is designed for the humidity
measurements in pressurized rooms and industrial processes. This
probe is also available for dewpoint measurements in natural gas. The
probe is provided with a nut, a fitting screw and a sealing washer to
make the installation easier.
1. duct wall
2. plugged hole for
reference measurements
is recommended
3. flange
4. seal
5. supporting bar
6. sensor head
1. a plugged
hole for
reference
measurements
2. sensor head
3. mounting
flange