Operation _________________________________________________________________________
Vaisala _________________________________________________________________________ 31
p
and
q
CALIBRATION RH
,
T
,
ANALOG
:
with these combinations you can perform relative humidity,
temperature or analog output calibrations with a multimeter or with
the transmitter display unit. Turn the dip switches to the desired
position according to the table printed on the protective cover.
r
OUTPUT QUANTITIES
:
determines whether the output units are metric (down) on non-metric.
s
,
t
and
u
SELECT OUTPUT QUANTITIES
:
with the three dip switches on the right, you can select the output
quantities according to the table printed on the right side of the
protective cover. Note that only the ordered quantities can be selected.
The
SPECIAL
option gives you the choice of setting any ordered
quantity to each channel.
NOTE
Always restore the dip switch settings after having tested the analog
outputs or performed the calibration.