Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 43
Stopping the continuous output
S
↵
The continuous outputting is stopped with the command
S
. If
outputting is active, this command is not echoed.
Setting the output interval
INTV n xxx
↵
where:
n
= 1 - 255
xxx
=
S
,
MIN
or
H
Sets the output interval when the transmitter outputs measurement
readings. The time interval is used when the continuous output is
active. For example, the output interval is set to 10 minutes:
>INTV 10 min
↵
Output intrv. :
10 min
>
Setting the ambient pressure for calculations
Turn the internal dip switch
CALIBRATION ENABLED/DISABLED
to
ON-position before setting the pressure and return it to OFF- position
after setting.
PRES aaaa.a
↵
where: aaaa.a = pressure (hPa)
>
PRES 2400
↵
Pressure : 2400
>
Pressure conversion chart
mmHg/Torr
inHg
atm
bar
psi
hPa/mbar
1
1.333224
33.86388
1013.25
1000
68.94757
mmHg/Torr
0.7500617
1
25.40000
760
750.0617
51.71493
inHg
0.02952999
0.03937008
1
29.921
29.52999
2.036021
atm
0.00098692
0.00131597
0.033422
1
0.98692
0.068046
bar
0.001
0.001333224
0.03386388
1.01325
1
0.06894757
psi
0.01450377
0.01933678
0.4911541
14.6962
14.50377
1
Example
: 29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa / mbar
NOTE
Conversions from mmHg and inHg are defined at 0 °C and for
mmH
2
O and inH
2
O at 4 °C.
FROM
TO