Error message
Description
Recommended action
Software restart
Device has automatically
restarted itself
Check that supply voltage is
stable and operating
environment is within
specification
Unexpected device restart
Status messages
Calibration is about to expire
Shown by calibration reminder
functionality
Calibrate the probe and update
the calibration date information
Purge in progress
Chemical purge of the humidity
sensor ongoing
Wait for chemical purge to
complete
The readings stay frozen until
the sensor cools down
Probe is waiting for its sensors
to cool down
Wait for measurement readings
to become available