Compensation setpoints
821
0334
hex
Temperature
compensation setpoint. If
a value is written to this
register, probe uses it
instead of its own
temperature
measurement.
When probe heating is in
use, temperature must be
written to this register to
enable output parameters
that depend on
temperature measurement
(for example, relative
humidity).
32-bit float
Unit: °C
0335
hex
Purge
773
0304
hex
Purge interval
32-bit float
Unit: min
10 ... 14400
0305
hex
1283
0502
hex
Interval purge on/off
16-bit
boolean
0 = Off
1 = On
1284
0503
hex
Startup purge on/off
16-bit
boolean
0 = Off
1 = On
Filtering
795
031A
hex
Measurement filtering
factor
32-bit float
Range: 0.000 ... 1.000
1.000 = Reading shows
100% of the most recent
measured value (no
filtering, default)
0.01 ... 0.99 = Reading
shows 1 ... 99% of the most
recent measured value and
part of the previous
reading (filtering is
applied). For example,
"0.9" means that the
filtered measurement
reading = 90% of the most
recent measured value +
10% of the previous
reading.
031B
hex
Appendix A – Modbus reference
83
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:[email protected]
Содержание HMP3
Страница 94: ...陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...
Страница 95: ...陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...
Страница 96: ...www vaisala com 陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...