62
20901101_EN_PC5xx-6xx_VARIO_select_Serie_V1.0_280618
Cleaning and maintenance
Example
Prepare
maintenance
7.3.2 Change the diaphragms and valves
Preparation
1
2
1.
Switch the Pumping unit off
and unplug the power plug.
2.
Remove the glass flask from
the EK as well as the connect-
ed hoses.
3
4
3.
Unscrew the screws from the
counterholder;
Torx screwdriver TX10
4.
Remove the counterholder and
lay it aside together with the
screws.
5
6
5.
Unscrew the knurled nuts,
pull off the molded hose, and
remove the cooler.
6.
Set the chiller down securely
so that no liquid can escape.
Example
Disassemble EK
(optionally)