55
20901101_EN_PC5xx-6xx_VARIO_select_Serie_V1.0_280618
Cleaning and maintenance
7 Cleaning and maintenance
WARNING
Danger due to electrical voltage.
>
>
Switch the device off before cleaning or maintenance
work.
>
>
Unplug the power plug from the socket.
Danger due to outdated motor condensers.
Condensers can become hot and possibly melt. There
can also rarely be darting flames.
>
>
Regularly check motor condensers (measure capac
-
ity, estimate operating hours). Typical service life is
10,000 to 40,000 operating hours, depending on op
-
erating conditions (temperature, engine load, ...).
>
>
Condensers must be replaced by an electrician.
Risk from contaminated parts.
Pumping hazardous media can result in hazardous
substances adhering to internal parts of the pump.
>
>
Wear your personal protective equipment, e. g., pro
-
tective gloves, eye protection and, if necessary, re
-
spiratory protection.
>
>
Decontaminate the vacuum pump before opening it.
If necessary have decontamination carried out by an
external service provider.
>
>
Take safety precautions according to your instruc
-
tions for handling hazardous substances.
NOTICE
Damage possible if work is performed incorrectly.
>
>
Have maintenance work performed by a trained professional
or at least by a trained person.
>
>
Recommendation: Before carrying out maintenance for the
first time, please read through all the instructions to get an
overview of the required service work.
Содержание PC 510 select
Страница 24: ...25 20901101_EN_PC5xx 6xx_VARIO_select_Serie_V1 0_280618 Safety information...
Страница 85: ...86 20901101_EN_PC5xx 6xx_VARIO_select_Serie_V1 0_280618...
Страница 86: ...87 20901101_EN_PC5xx 6xx_VARIO_select_Serie_V1 0_280618...
Страница 87: ...88 20901101_EN_PC5xx 6xx_VARIO_select_Serie_V1 0_280618...