Program
N0
hh:mm:ss
Vac Vent. Step
01
00:00:00
ATM
✔
✔
02
00:05:00
10
03
00:15:00
10
04
00:01:00
500
✔
05
00:09:00
500
✔
06
00:10:00
5
07
00:20:00
5
08
00:01:00
ATM
✔
✔
09
00:00:00
10
00:00:00
- - - - - - - - - - - - - Back - - - - - - - - - - - - -
01:01:00
Program step 1 should be always a definite initial state, here atmospheric pressure. To reach this state
definitely set a tickmark at venting and ”Step” (press selection knob).
Step 2 causes pumping down to 10 mbar within 5 minutes (linear ramp) as ”Step” is not set.
Step 3 keeps the vacuum (10 mbar) constant for 15 minutes.
Step 4 vents within 1 minute to 500 mbar.
Step 5 keeps 500 mbar for 9 minutes.
Step 6 pumps down to 5 mbar within 10 minutes.
Step 7 keeps the vacuum (5 mbar) constant for 20 minutes.
Step 8 vents as fast as possible to atmospheric pressure and switches off the control after one minute.
Vacuum pump with isolation valve and/or Vacuum-Management-System Module A: Pumping
down with intermediate venting
Example for use
Once the program is finished, the clock symbol starts to flash. Confirm the end of the program by press
-
ing START/STOP (clock symbol will disappear).
Attention
: If
”Autostart”
is set to
”
On
”
, the program will start again (time will be reset to 00:00:00)
after a power failure or after switching the controller off/on. Only if the end of the program (clock symbol
flashing) has been confirmed by pressing START/STOP, the program will not start again.