Catchpots:
The catchpot at the inlet protects against drop-
lets and particles from entering the pump.
+
Enhances lifetimes of diaphragms and
valves.
+
Improves vacuum performance in applica-
tions with condensable vapors.
Both catchpots are coated with a protective
layer to protect against shattering in case of
breakage or implosion.
➨
Assemble the catchpots at the inlet and at
the outlet using joint clips.
Exhaust waste vapor condenser:
➨
Assemble the hose nozzles for coolant inlet
(12) and coolant outlet (13) tubing at the ex-
haust waste vapor condenser.
The exhaust waste vapor condenser enables
an efficient condensation of the pumped va
-
pors at the outlet.
+
No backflow of condensates.
+
Controlled recovery of condensates.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass splinters in
case of breakage, acts as thermal isolation to avoid
condensation of humidity and is intended to absorb
shocks.
➨
Attach the tubing of the coolant circuit to the respective
hose nozzles (hose nozzles for tubing I.D. 1/4”-5/16”
(6-8 mm), see image) at the waste vapor condenser.
Check all hose connections prior to starting operation of
the cooling system.
Secure coolant hoses at the hose nozzles (e.g., with hose
clamps) to prevent their accidentally slipping off.
catchpot at
the outlet
catchpot at
the inlet
(9)
outlet (gas!)
2
12
13
NOTICE