![vacuubrand PC 510 NT Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/vacuubrand/pc-510-nt/pc-510-nt_instructions-for-use-manual_3191774030.webp)
page 30 of 118
Gas inlet temperatures
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
Operating condition
Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨
50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨
32 °F to 140 °F*
(0°C to +60°C*)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨
14 °F to 176 °F*
(-10°C to +80°C*)