Installation and connection
20901552_EN-US_PC301x NT V.sel (Ser.Atb_Ha)_V1.1_160822
29
Installing the vacuum pump
-> Example
Sketch of minimum
distances in labora-
tory furniture
ð
Place the vacuum pump on a stable, vibration-free, level sur-
face.
ð
When installing laboratory furniture, maintain the minimum
distance of 5 cm (2 in.) to adjacent objects or surfaces.
ð
Avoid heat accumulation and ensure sufficient air circulation,
especially in closed housings.
Observing the limitations of use
Ambient conditions
Ambient conditions
(US)
Ambient temperature
10–40 °C
50–104°F
Installation height, max.
2000 m
above sea level
6562 ft
above sea level
Humidity
30–85 %, non-condensing
Degree of contamination
2
Impact energy
5 J
Protection class (IEC 60529)
IP 40
Protection class (UL 50E)
Type 1
Avoid condensate or contamination from dust, liquids and corrosive gases.
ð
Note the indicated IP protection rating. The IP protection is only
guaranteed if the device properly installed and connected.
ð
When connecting the device, always take note of the specifica-
tions from the type plate and in the chapter technical data.