Troubleshooting
52
20901334_EN-US_PC3000 VARIO select Serie_V1.4_061022
Error
Cause
Corrective measure Personnel
Check the gas ballast
cap.
Replace defective
components.
No suction capacity
or very little
Deposits in the vac-
uum pump.
Membrane or valves
defective.
High amount of va-
por build up in
process.
Clean and test pump
heads.
Replace the mem-
brane and valves.
Check process pa-
rameters.
Qualified
employee
No suction capacity
or very little
Vacuum line too
long.
Use vacuum lines
with a larger cross-
section.
Resp. spe-
cialist
Display off
Pumping unit
switched off.
Power plug not con-
nected properly or
pulled out.
VACUU BUS plug
connection or cable
defective or not in-
serted.
Controller switched
off or defective.
Switch on Pumping
unit.
Check the power
connection and ca-
ble.
Check the VACUU
BUS plug connection
and cable to the con-
troller.
Replace defective
components.
Operator
Condenser (cooler)
defective
Mechanically dam-
aged.
Send in.
Resp. spe-
cialist
Loud operating
noises
No hose installed.
Check hose and in-
stall correctly.
Operator
Loud operating
noises
Exhaust line open.
Glass flask on EK
missing.
Membrane torn or
membrane spring
washer loose.
Ball bearings defec-
tive.
Check exhaust line
connections.
Connect exhaust line
to a suction or ex-
traction system.
Install glass flask
Service vacuum
pump and replace
defective parts, or
send device in for re-
pair.
Qualified
employee
Vacuum pump
stopped and will not
start again.
Speed setting too
low for motor type
and application con-
figuration.
Set the speed to >
15% in the con-
troller.
Operator
Содержание PC 3002 VARIO select
Страница 82: ...Annex 82 20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 8 6 EU declaration of conformity ...
Страница 83: ...Annex 20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 83 8 7 UKCA conformity declaration ...
Страница 86: ...86 20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 ...
Страница 87: ...20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 87 ...