Troubleshooting
50
20901334_EN-US_PC3000 VARIO select Serie_V1.4_061022
6
Troubleshooting
6.1 Technical assistance
To find and eliminate errors, use the table
For technical assistance or in case of faults, please contact our
department.
The device should only be operated in perfect technical condition.
ð
Adhere to the recommended maintenance intervals to ensure a
properly functioning system.
ð
Send defective devices to our Service department or your spe-
cialist retailer for repairs.
6.2 Error – cause – corrective measure
Error
Cause
Corrective measure Personnel
Measurements devi-
ate from the refer-
ence standard
Sensor dirty.
Humidity in sensor.
Sensor defective.
Sensor not measur-
ing correctly.
Clean sensor mea-
suring chamber.
Allow sensor measur-
ing chamber to dry,
e.g. by draining.
Calibrate sensor with
reference measuring
device.
Replace defective
components.
Qualified
employee
Sensor does not
transmit measure-
ment
No voltage applied.
VACUU BUS plug
connection or cable
defective or not in-
serted.
Check the VACUU
BUS plug connection
and cable to the con-
troller.
Operator
Sensor does not
transmit measure-
ment
Sensor defective.
Replace defective
components.
Qualified
employee
Ventilation valve
does not switch
No voltage applied.
VACUU BUS plug
connection or cable
defective or not in-
serted.
Ventilation valve
dirty.
Check the VACUU
BUS plug connection
and cable to the con-
troller.
Clean the ventilation
valve.
Operator
Содержание PC 3002 VARIO select
Страница 82: ...Annex 82 20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 8 6 EU declaration of conformity ...
Страница 83: ...Annex 20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 83 8 7 UKCA conformity declaration ...
Страница 86: ...86 20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 ...
Страница 87: ...20901334_EN US_PC3000 VARIO select Serie_V1 4_061022 87 ...