background image

page 8 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully 

the validity of this document with respect to his product.  Manual-no.:  999198 / 29/09/2009

We reserve the right for technical modifications without prior note!

Technical data

Information: In a motor speed range between

 0 and 200 rpm

 the pump is running automatically in a 

cycled interval operation.

Type

PC 3001 basic

Maximum pumping speed  (ISO 21360)

m

3

/h

1.7

Ultimate vacuum (absolute) without gas ballast

mbar

2

Ultimate vacuum (absolute) with gas ballast

mbar

4

Maximum permissible inlet pressure (absolute)

bar

1.1

Maximum permissible outlet pressure (abso-

lute)

bar

1.1

Maximum permissible pressure (absolute) at 

gas ballast valve

bar

1.2

Permissible ambient temperature storage / op-

eration

°C

-10 to +60 / +10 to +40

Permissible relative atmospheric moisture dur-

ing operation (no condensation)

%

30 to 85

Rated power (electric)

kW

0.16

No-load speed*

min

-1

0 - 2400

Maximum permissible range of supply voltage 

±

10% ) 

Attention: Observe specifications 

of rating plate!

100-120 V~ 50-60 Hz 

200-230 V~ 50-60 Hz

Maximum rated current at:

 100-120 V~ 50/60 Hz

 200-230 V~ 50/60 Hz

A

A

1.6

0.7

Device fuse

1x 7A 

1x 0.5A self-resetting (

VACUU•BUS)

Motor protection

temperature sensor on the pcb

(current limitation)

Degree of protection IEC 529

IP 20

Inlet

hose nozzle DN 6/10 mm

Outlet

hose nozzle DN 10 mm / silencer

Dimensions L x W x H  approx.

mm

251 x 256 x 400

Weight approx.

kg

6.4

Содержание PC 3001 basic

Страница 1: ...y and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999198 29 09 2009 PC 3001 basic Chemistry pumping unit wi...

Страница 2: ...powerful application usage and personal safety through our products Please read these instructions for use before the initial operation of your pump VACUUBRAND diaphragm pumps are the result of many y...

Страница 3: ...inlet temperatures 9 Wetted parts 9 Pump parts 10 Kits for PC 3001 basic 11 Use and operation 12 Installing in a vacuum system 12 During operation 13 Attention Important notes regarding the use of gas...

Страница 4: ...ion between 10 C and 40 C Check the maximum temperatures if installing the pump in a cabinet or a housing and make sure ventilation is adequate Install an external automatic ventilation system if nece...

Страница 5: ...nt into a room for operation allow the equipment to warm up pay attention to water condensation on cold surfaces The diameter of the inlet and outlet pipeline should be at least as large as the diame...

Страница 6: ...t of the human body from coming into contact with vacuum The user must take suitable precautions to prevent any formation of explosive mixtures in the expansion chamber or at the outlet In case of e g...

Страница 7: ...t maximum Prevent any backpressure of gases and the backflow of condensates Never suck liquids or dust into the pump The A weighed emission sound pressure level of the pump does not exceed 70 dB A Mea...

Страница 8: ...e bar 1 1 Maximum permissible outlet pressure abso lute bar 1 1 Maximum permissible pressure absolute at gas ballast valve bar 1 2 Permissible ambient temperature storage op eration C 10 to 60 10 to 4...

Страница 9: ...ssure Permitted range of gas temperatures at inlet Continuous operation 100 mbar high gas load 10 C to 40 C Continuous operation 100 mbar low gas load 0 C to 60 C Short time 5 minutes 100 mbar low gas...

Страница 10: ...the validity of this document with respect to his product Manual no 999198 29 09 2009 PC 3001 basic Position Component 1 Chemistry diaphragm pump MD 1C VARIO 2 knob to select motor speed 3 Mains conn...

Страница 11: ...knob to the right increases the motor speed A high motor speed reduces the lifetime of diaphragms and valves Pumping at high motor speed increases the pumping speed of the pump Ensure sufficient cooli...

Страница 12: ...tack and deposits replace if necessary or use an exhaust pipeline without silencer Reduce the transmission of vibration and prevent loading due to rigid pipelines Insert elastic hoses or flexible elem...

Страница 13: ...ust The diaphragm and valves will be damaged if liquids are pumped in significant amounts Check the pump regularly for external soiling and deposits clean if necessary to avoid an increase of the pump...

Страница 14: ...rms deposits Check and clean pump heads if necessary Long term Take measures as described in section short term shutdown Separate pump from the application Close gas ballast Drain catchpots if using w...

Страница 15: ...ultimate vacuum or nor mal pumping speed Centring ring at a small flange connection not correctly posi tioned or leak in the pipeline or vacuum system Check pump directly con nect vacuum gauge direct...

Страница 16: ...999198 29 09 2009 Kits Kit AK CVC 3000 699921 Kit EK 500 699922 Accessories for vacuum controller CVC 3000 External pressure transducer VSK 3000 capacitive ceramic diaphragm sensor 1080 0 1 mbar 63665...

Страница 17: ...ensation transfer of liquids or dust The diaphragm and valves will be damaged if liquids are pumped in significant amount If the pump is exposed to corrosive gases or vapour or in case of deposits mai...

Страница 18: ...from one side of the housing cover Disassembling the housing cover at the inlet side Lay the pumping unit down with the housing cover at the side of the pump inlet facing upwards support pumping unit...

Страница 19: ...ent with respect to his product Manual no 999198 29 09 2009 View of the disassembled pump head parts Replacing the diaphragm 10 1 9 3 2 5 6 7 8 4 1 housing cover 2 valve 3 head cover 4 diaphragm clamp...

Страница 20: ...ply pressure to the diaphragm clamping disc to bring the diaphragm into the lower turning point position Press diaphragm key against diaphragm clamping disc unscrew and remove diaphragm support disc w...

Страница 21: ...side with hose connection between pump heads side with knob stand foot Assembling the housing cover at the inlet side Assemble head covers and valves Check for correct position see also fig below It...

Страница 22: ...p ing unit appropriately Use a hex key size 4 to loosen the six socket head screws at the pump head Raise the housing cover Pull off the hose from the housing cover Do not kink the hose Remove head co...

Страница 23: ...nge after the run in period Check hose connectors between pump heads and manifolds for leaks If necessary recheck pump chamber Position housing cover Pay attention to correct hose guid ance Move housi...

Страница 24: ...onditions with every product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair...

Страница 25: ...oxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oil or residues of pumped media have been...

Страница 26: ...gn ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration cesse d tre valable...

Страница 27: ...obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of app...

Отзывы: