background image

page 7 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully 

the validity of this document with respect to his product. Manual-no.:   999198 / 29/09/2009

•  Due to the residual 

leak rate of the equipment

, there might be an exchange of 

gas, albeit extremely slight, between the environment and the vacuum system.

  Adopt suitable measures to prevent contamination of the pumped substances or 

the environment.

Provide appropriate protective measures (i.e precautions which allow for the require-

ments of the respective application) even for the case of failure and 

malfunction

.

Failure of the pump (e.g. due to power failure) or of connected components, of parts 

of the supply or change of parameters must not lead to a critical dangerous situa-

tion under any circumstances. In case of diaphragm cracks or leaks in the manifold 

pumped substances might be released into the environment or into the pump hous-

ing or motor. Comply especially with notes on operation and use and maintenance. 

The  motor  is  protected  by  a

  temperature  sensor  at  the  circuit  board: 

Current 

limitation if the temperature at the circuit board is higher than 70°C, switching off the 

pump if the temperature is higher than 85°C. In case of blockade of the motor (after 

10 attempts to start-up) the pump is switched off.

If the pump is switched off due to safety measures, manual reset is necessary. Isolate 

the pump from mains. Eliminate the cause of failure before restarting the pump.

Do not start the pump if the pressure difference between inlet and outlet port ex-

ceeds 1.1 bar at maximum.

Prevent any backpressure of gases and the backflow of condensates.

Never suck liquids or dust into the pump.

The A-weighed  emission  sound  pressure  level  of  the  pump  does  not  exceed  70 

dB(A). Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 with 

standard silencer or exhaust tube at outlet.

Maintenance and repair

Wear parts have to be replaced regularly. 

In case of normal wear the lifetime of 

the diaphragms and valves is > 10000 operating hours. Bearings have a typical du-

rability of 40000 h. 

 

Isolate equipment from mains

 and 

wait two minutes

 before starting mainte-

nance to allow the capacitors to discharge. 

  Ensure that the pump cannot be operated accidentally. Never operate the pump 

if covers or other parts of the pump are disassembled. Never operate a defective 

or damaged pump.

 

Attention

: The pump might be contaminated with process chemicals which have 

been pumped during operation. Ensure that the pump is decontaminated before 

maintenance and take adequate precautions to protect people from the effects of 

dangerous substances if contamination has occurred.

•  Before starting maintenance vent the pump, isolate the pump and other compo

-

nents from the vacuum system. Allow sufficient cooling of the pump. Drain con

-

densate, if applicable.

Ensure that 

maintenance

 is done only by suitably trained and supervised techni-

cians. Ensure that the maintenance technician is familiar with the safety procedures 

which relate to the products processed by the pumping system. 

In order to comply with law (occupational, health and safety regulations, safety at 

work law and regulations for environmental protection) vacuum pumps, components 

and measuring instruments returned to the manufacturer can be repaired only when 

certain procedures (see section 

”Notes on return to the factory”

) are followed.

NOTICE

Содержание PC 3001 basic

Страница 1: ...y and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999198 29 09 2009 PC 3001 basic Chemistry pumping unit wi...

Страница 2: ...powerful application usage and personal safety through our products Please read these instructions for use before the initial operation of your pump VACUUBRAND diaphragm pumps are the result of many y...

Страница 3: ...inlet temperatures 9 Wetted parts 9 Pump parts 10 Kits for PC 3001 basic 11 Use and operation 12 Installing in a vacuum system 12 During operation 13 Attention Important notes regarding the use of gas...

Страница 4: ...ion between 10 C and 40 C Check the maximum temperatures if installing the pump in a cabinet or a housing and make sure ventilation is adequate Install an external automatic ventilation system if nece...

Страница 5: ...nt into a room for operation allow the equipment to warm up pay attention to water condensation on cold surfaces The diameter of the inlet and outlet pipeline should be at least as large as the diame...

Страница 6: ...t of the human body from coming into contact with vacuum The user must take suitable precautions to prevent any formation of explosive mixtures in the expansion chamber or at the outlet In case of e g...

Страница 7: ...t maximum Prevent any backpressure of gases and the backflow of condensates Never suck liquids or dust into the pump The A weighed emission sound pressure level of the pump does not exceed 70 dB A Mea...

Страница 8: ...e bar 1 1 Maximum permissible outlet pressure abso lute bar 1 1 Maximum permissible pressure absolute at gas ballast valve bar 1 2 Permissible ambient temperature storage op eration C 10 to 60 10 to 4...

Страница 9: ...ssure Permitted range of gas temperatures at inlet Continuous operation 100 mbar high gas load 10 C to 40 C Continuous operation 100 mbar low gas load 0 C to 60 C Short time 5 minutes 100 mbar low gas...

Страница 10: ...the validity of this document with respect to his product Manual no 999198 29 09 2009 PC 3001 basic Position Component 1 Chemistry diaphragm pump MD 1C VARIO 2 knob to select motor speed 3 Mains conn...

Страница 11: ...knob to the right increases the motor speed A high motor speed reduces the lifetime of diaphragms and valves Pumping at high motor speed increases the pumping speed of the pump Ensure sufficient cooli...

Страница 12: ...tack and deposits replace if necessary or use an exhaust pipeline without silencer Reduce the transmission of vibration and prevent loading due to rigid pipelines Insert elastic hoses or flexible elem...

Страница 13: ...ust The diaphragm and valves will be damaged if liquids are pumped in significant amounts Check the pump regularly for external soiling and deposits clean if necessary to avoid an increase of the pump...

Страница 14: ...rms deposits Check and clean pump heads if necessary Long term Take measures as described in section short term shutdown Separate pump from the application Close gas ballast Drain catchpots if using w...

Страница 15: ...ultimate vacuum or nor mal pumping speed Centring ring at a small flange connection not correctly posi tioned or leak in the pipeline or vacuum system Check pump directly con nect vacuum gauge direct...

Страница 16: ...999198 29 09 2009 Kits Kit AK CVC 3000 699921 Kit EK 500 699922 Accessories for vacuum controller CVC 3000 External pressure transducer VSK 3000 capacitive ceramic diaphragm sensor 1080 0 1 mbar 63665...

Страница 17: ...ensation transfer of liquids or dust The diaphragm and valves will be damaged if liquids are pumped in significant amount If the pump is exposed to corrosive gases or vapour or in case of deposits mai...

Страница 18: ...from one side of the housing cover Disassembling the housing cover at the inlet side Lay the pumping unit down with the housing cover at the side of the pump inlet facing upwards support pumping unit...

Страница 19: ...ent with respect to his product Manual no 999198 29 09 2009 View of the disassembled pump head parts Replacing the diaphragm 10 1 9 3 2 5 6 7 8 4 1 housing cover 2 valve 3 head cover 4 diaphragm clamp...

Страница 20: ...ply pressure to the diaphragm clamping disc to bring the diaphragm into the lower turning point position Press diaphragm key against diaphragm clamping disc unscrew and remove diaphragm support disc w...

Страница 21: ...side with hose connection between pump heads side with knob stand foot Assembling the housing cover at the inlet side Assemble head covers and valves Check for correct position see also fig below It...

Страница 22: ...p ing unit appropriately Use a hex key size 4 to loosen the six socket head screws at the pump head Raise the housing cover Pull off the hose from the housing cover Do not kink the hose Remove head co...

Страница 23: ...nge after the run in period Check hose connectors between pump heads and manifolds for leaks If necessary recheck pump chamber Position housing cover Pay attention to correct hose guid ance Move housi...

Страница 24: ...onditions with every product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair...

Страница 25: ...oxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oil or residues of pumped media have been...

Страница 26: ...gn ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration cesse d tre valable...

Страница 27: ...obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of app...

Отзывы: