Troubleshooting

Fault

Possible cause

Remedy

Pump does not start or 

stops immediately.

Knob for motor speed is 

turned to the left stop?

Turn the knob to the right.

Supply voltage is missing or 

too low?

Check or connect the supply 

voltage; check the mains 

fuse.

Overpressure in the exhaust 

line?

Open exhaust line, ensure a 

free (pressureless) outlet line

Motor overloaded?

Allow motor to cool. Identify 

and eliminate the cause of 

failure. 

VACUU•BUS fuse has been 

triggered, e.g. load in the 

VACUU•BUS system too high 

or short circuit in valves?

Remove VACUU•BUS com

-

ponents (e.g. valves). The 

fuse is self-resetting.

Pump does not achieve 

ultimate vacuum or nor-

mal pumping speed.

Centring ring at a small flange 

connection not correctly posi-

tioned or leak in the pipeline 

or vacuum system?

Check pump directly - con-

nect vacuum gauge directly 

at pump inlet - then check 

connection, pipeline and 

vacuum system if necessary.

Long, narrow vacuum line?

Use lines with larger dia-

meter, length as short as 

possible.

Pump has been exposed to 

condensate?

Allow pump to run for some 

minutes with atmospheric 

pressure at the inlet.

Deposits have been formed 

inside the pump?

Clean and inspect the pump 

heads.

Diaphragms or valves defec-

tive?

Replace diaphragms and/or 

valves.

Outgassing substances or 

vapour generated in the pro-

cess?

Check process parameters.

Pump temperature too high 

(reduced motor speed)?

Ensure sufficient cooling 

of the pump or reduce inlet 

pressure.

Pump too noisy.

Loud exhaust noise?

Connect hose or silencer to 

pump outlet.

Diaphragm crack or dia-

phragm clamping disc loose?

Perform maintenance.

Other than above mentioned 

causes?

Contact local distributor.

Pump seized.

Contact local distributor.

Содержание PC 3001 basic

Страница 1: ...y and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999198 29 09 2009 PC 3001 basic Chemistry pumping unit wi...

Страница 2: ...powerful application usage and personal safety through our products Please read these instructions for use before the initial operation of your pump VACUUBRAND diaphragm pumps are the result of many y...

Страница 3: ...inlet temperatures 9 Wetted parts 9 Pump parts 10 Kits for PC 3001 basic 11 Use and operation 12 Installing in a vacuum system 12 During operation 13 Attention Important notes regarding the use of gas...

Страница 4: ...ion between 10 C and 40 C Check the maximum temperatures if installing the pump in a cabinet or a housing and make sure ventilation is adequate Install an external automatic ventilation system if nece...

Страница 5: ...nt into a room for operation allow the equipment to warm up pay attention to water condensation on cold surfaces The diameter of the inlet and outlet pipeline should be at least as large as the diame...

Страница 6: ...t of the human body from coming into contact with vacuum The user must take suitable precautions to prevent any formation of explosive mixtures in the expansion chamber or at the outlet In case of e g...

Страница 7: ...t maximum Prevent any backpressure of gases and the backflow of condensates Never suck liquids or dust into the pump The A weighed emission sound pressure level of the pump does not exceed 70 dB A Mea...

Страница 8: ...e bar 1 1 Maximum permissible outlet pressure abso lute bar 1 1 Maximum permissible pressure absolute at gas ballast valve bar 1 2 Permissible ambient temperature storage op eration C 10 to 60 10 to 4...

Страница 9: ...ssure Permitted range of gas temperatures at inlet Continuous operation 100 mbar high gas load 10 C to 40 C Continuous operation 100 mbar low gas load 0 C to 60 C Short time 5 minutes 100 mbar low gas...

Страница 10: ...the validity of this document with respect to his product Manual no 999198 29 09 2009 PC 3001 basic Position Component 1 Chemistry diaphragm pump MD 1C VARIO 2 knob to select motor speed 3 Mains conn...

Страница 11: ...knob to the right increases the motor speed A high motor speed reduces the lifetime of diaphragms and valves Pumping at high motor speed increases the pumping speed of the pump Ensure sufficient cooli...

Страница 12: ...tack and deposits replace if necessary or use an exhaust pipeline without silencer Reduce the transmission of vibration and prevent loading due to rigid pipelines Insert elastic hoses or flexible elem...

Страница 13: ...ust The diaphragm and valves will be damaged if liquids are pumped in significant amounts Check the pump regularly for external soiling and deposits clean if necessary to avoid an increase of the pump...

Страница 14: ...rms deposits Check and clean pump heads if necessary Long term Take measures as described in section short term shutdown Separate pump from the application Close gas ballast Drain catchpots if using w...

Страница 15: ...ultimate vacuum or nor mal pumping speed Centring ring at a small flange connection not correctly posi tioned or leak in the pipeline or vacuum system Check pump directly con nect vacuum gauge direct...

Страница 16: ...999198 29 09 2009 Kits Kit AK CVC 3000 699921 Kit EK 500 699922 Accessories for vacuum controller CVC 3000 External pressure transducer VSK 3000 capacitive ceramic diaphragm sensor 1080 0 1 mbar 63665...

Страница 17: ...ensation transfer of liquids or dust The diaphragm and valves will be damaged if liquids are pumped in significant amount If the pump is exposed to corrosive gases or vapour or in case of deposits mai...

Страница 18: ...from one side of the housing cover Disassembling the housing cover at the inlet side Lay the pumping unit down with the housing cover at the side of the pump inlet facing upwards support pumping unit...

Страница 19: ...ent with respect to his product Manual no 999198 29 09 2009 View of the disassembled pump head parts Replacing the diaphragm 10 1 9 3 2 5 6 7 8 4 1 housing cover 2 valve 3 head cover 4 diaphragm clamp...

Страница 20: ...ply pressure to the diaphragm clamping disc to bring the diaphragm into the lower turning point position Press diaphragm key against diaphragm clamping disc unscrew and remove diaphragm support disc w...

Страница 21: ...side with hose connection between pump heads side with knob stand foot Assembling the housing cover at the inlet side Assemble head covers and valves Check for correct position see also fig below It...

Страница 22: ...p ing unit appropriately Use a hex key size 4 to loosen the six socket head screws at the pump head Raise the housing cover Pull off the hose from the housing cover Do not kink the hose Remove head co...

Страница 23: ...nge after the run in period Check hose connectors between pump heads and manifolds for leaks If necessary recheck pump chamber Position housing cover Pay attention to correct hose guid ance Move housi...

Страница 24: ...onditions with every product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair...

Страница 25: ...oxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oil or residues of pumped media have been...

Страница 26: ...gn ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration cesse d tre valable...

Страница 27: ...obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of app...

Отзывы: