![vacuubrand MZ 2 VARIO-SP Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/vacuubrand/mz-2-vario-sp/mz-2-vario-sp_instructions-for-use-manual_3211332037.webp)
Fault
Possible cause
Remedy
❑
Pump too noisy.
➨
Atmospheric or high
pressure at the pump
inlet?
✔
Connect hose or silencer
to pump outlet. Be careful
not to cause outlet over-
pressure, especially with
condensable vapors.
➨
Diaphragm crack or
diaphragm clamping
disc loose?
✔
Perform maintenance.
➨
Other than above men-
tioned causes?
✔
Contact local distributor.
❑
Pump seized.
✔
Contact local distributor.
➨
A service manual with exploded view drawings, spare parts list and
directions for repair is available on request.
+
The service manual is intended for trained service people only.