Adopt suitable measures to prevent any danger aris-
ing from hot surfaces or electric sparks. Ensure that
hot surfaces of the pump do not cause burns. Provide
a suitable contact guard if necessary.
Prevent the backpressure of gases and the backflow of
condensates at the outlet.
Provide appropriate protective measures to allow for the
possibility of failure and
malfunction
. The protective mea-
sures must also allow for the requirements of the respec-
tive application.
A temperature sensor at the circuit board protects the mo-
tor: Current limitation in case the temperature at the circuit
board raises above 158°F (70°C). At very high tempera-
tures the pump is switched off.
Note
: Only manual reset is possible. Disconnect the pump
from the power source. Identify and eliminate the cause of
failure.
Maintenance and repair
In order to comply with laws (occupational, health and
safety regulations, safety at work law and regulations for
environmental protection) vacuum pumps, components
and measuring instruments can only be returned when
certain procedures (see section „
Take advantage of our service seminars, which put special
focus on the maintenance and repair of vacuum pumps.
For details and for the online ”Instructions for repair” man-
In normal use, the lifetime of the diaphragms and valves
is typically 15,000 operating hours (at 1500 rpm). Bear-
ings have a typical durability of 40000 h.