Installing a pump in a vacuum system
➨
If dangerous or polluting fluids could be released at
the outlet, install an appropriate system to catch and
dispose of those fluids.
+
Connect a gas-tight exhaust line at the pump outlet if
necessary. Always vent exhaust gases appropriately
(e.g., into a fume hood).
+
Never block or restrict the gas outlet. The exhaust line
must always be free of obstructions (no back pressure)
to ensure an unimpeded discharge of gas. The cross-
section of the outlet tubing must be at least the size of
the pump’s exhaust connection.
+
Particles and dust must not enter the pump. If nec-
essary, you must install appropriate filters. You must
ensure their suitability concerning gas flow, chemical
resistance and resistance to clogging prior to use.
+
Make sure ventilation is adequate, especially if the pump
is installed in an enclosure, or if the ambient tempera-
ture is elevated. Provide external ventilation, if neces-
sary.
• Reduce the transmission of vibration. Prevent mechan
-
ical load due to rigid pipelines. Insert elastic hoses or
flexible elements as couplings between the pump and
rigid pipes.
Note
: Flexible elements will compress or flatten when
evacuated if not designed for use under vacuum.
• Hose connections at the pump inlet must always be
gas tight.
• A power failure may cause accidental ventilation of the
pump. If this constitutes a potential source of danger,
take appropriate safety measures.
• Check the power source and the pump’s rating plate
to be sure that the power source and the equipment
match in voltage, phase, and frequency.